Case Study: PRESTO W55 Process system
Cooling a Büchiglas 100 l reactor from +20 °C to -20 °C with a PRESTO W55 Process system
Objective
This case study tests the cooling capacity of the JULABO PRESTO W55 Process system with a 100 l glass reactor. The PRESTO W55 is connected to the reactor via two 2 m long metal hoses. The PRESTO W55 Process system has been programmed to cool down from +20 °C to -20 °C.
Test Conditions
| Product | PRESTO W55 Process system | ||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||||||
| Heating capacity | 15 kW | ||||||||||||||||||
| Pump pressure | 0.5 bar | Band limit | °k | ||||||||||||||||
| Duration | 1:10 hr:min | Test volume | 75 l | ||||||||||||||||
| Setpoint 1 | +20 °C | Setpoint 2 | -20 °C | ||||||||||||||||
| Reactor | 100l Büchiglas (glass) | Tubing length | 2 m | ||||||||||||||||
| Fluid in unit | Thermal HL60 | Fluid in reactor | Thermal HL60 | ||||||||||||||||
Environment
| Room temperature | 20 °C |
|---|---|
| Humidity | 45 % |
| Voltage version | 400V/3PNPE/50Hz |
Control Parameters
| XP | 0.19 K |
| TN | 695 s |
| TV | 85 s |
| XPU | 15 K |
Test Results
The PRESTO W55 Process system cools the reactor from +20 °C to -20 °C in 1:10 h.
- Set point
- Temperature in the application
- Temperature in temperature control unit