FPW55-SL Ultra-Low Refrigerated-Heating Circulator
for working temperatures from -95 °C to +150 °C with filling port upgradeable with additional heating and pump capacity for external control only
Ultra-Low Refrigerated Circulators with SL circulator impress with their high heating, cooling and pump capacities for external temperature control applications.
Request a quote| Order No. | 9 352 756N.07 | |||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Working temperature range | -60 ... +100 °C | |||||||||||||
| Temperature stability | ±0.05 °C | |||||||||||||
| Heating capacity | 3 kW | |||||||||||||
| Cooling capacity (Medium: Ethanol) |
|
|
||||||||||||
|
Cooling capacity
(Medium: Ethanol) |
||||||||||||||
|
||||||||||||||
| Pump capacity flow rate | 22 ... 26 l/min | |||||||||||||
| Pump capacity flow pressure | 0.4 ... 0.7 bar | |||||||||||||
| Filling volume | 27 l | |||||||||||||
|
Dimensions (W × L × H) |
59 x 76 x 116 cm | |||||||||||||
| Order No. | 9 352 756N.16 | |||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Working temperature range | -60 ... +100 °C | |||||||||||||
| Temperature stability | ±0.05 °C | |||||||||||||
| Heating capacity | 3 kW | |||||||||||||
| Cooling capacity (Medium: Ethanol) |
|
|
||||||||||||
|
Cooling capacity
(Medium: Ethanol) |
||||||||||||||
|
||||||||||||||
| Pump capacity flow rate | 22 ... 26 l/min | |||||||||||||
| Pump capacity flow pressure | 0.4 ... 0.7 bar | |||||||||||||
| Filling volume | 27 l | |||||||||||||
|
Dimensions (W × L × H) |
59 x 76 x 116 cm | |||||||||||||
| Order No. | 9 352 756N.16 | |||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Working temperature range | -60 ... +100 °C | |||||||||||||
| Temperature stability | ±0.05 °C | |||||||||||||
| Heating capacity | 3 kW | |||||||||||||
| Cooling capacity (Medium: Ethanol) |
|
|
||||||||||||
|
Cooling capacity
(Medium: Ethanol) |
||||||||||||||
|
||||||||||||||
| Pump capacity flow rate | 22 ... 26 l/min | |||||||||||||
| Pump capacity flow pressure | 0.4 ... 0.7 bar | |||||||||||||
| Filling volume | 27 l | |||||||||||||
|
Dimensions (W × L × H) |
59 x 76 x 116 cm | |||||||||||||
Product Features
- VFD COMFORT DISPLAY
- LCD DIALOG DISPLAY backlit for convenient interactive operation
- Keypad for setpoints, warning/safety values and menu functions
- ICC (Intelligent Cascade Control), self-optimizing temperature control
- TCF Temperature Control Features to optimize the control behavior
- ATC3 3-Point-Calibration
- Pt100 External sensor connection for measurement and control
- SMART PUMP, electronically adjustable pump stages
- Adjustable high temperature cut-out, visible via display
- RS232/RS485 interface for online communication
- Integrated programmer for 6 x 60 program steps
- Connections for solenoid valve and HSP booster pump
- Proportional cooling control
|
Dimensions (W × L × H) |
59 x 76 x 116 cm |
|---|---|
| Filling volume | 27 l |
| Weight | 146 kg |
| Pump connections | M16x1 male |
| Cooling Water Connection | G¾ in |
| Working temperature range | -60 ... +100 °C |
|---|---|
| Temperature stability | ±0.05 °C |
| Setting the resolution of the temperature display | 0.01 °C |
| Temperature display resolution | 0.01 °C |
| Ambient temperature | +5.0 ... +40.0 °C |
| Heating capacity | 3 kW | ||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Cooling capacity (Medium: Ethanol) |
|
|
|||||||||||||
|
Cooling capacity
(Medium: Ethanol) |
|||||||||||||||
|
|||||||||||||||
| Pump capacity flow rate | 22 ... 26 l/min | ||||||||||||||
| Pump capacity flow pressure | 0.4 ... 0.7 bar | ||||||||||||||
| Maximum suction | 0.2 ... 0.4 bar | ||||||||||||||
| Viscosity max. | 70 cST | ||||||||||||||
Refrigerant stage 1 |
Refrigerant | R452A | |||||||||||||
| Global Warming Potential for R452A | 2140 | ||||||||||||||
| Filling volume | 850 g | ||||||||||||||
| Carbon dioxide equivalent | 1.819 t | ||||||||||||||
| Heating capacity | 3 kW | ||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Cooling capacity (Medium: Ethanol) |
|
|
|||||||||||||
|
Cooling capacity
(Medium: Ethanol) |
|||||||||||||||
|
|||||||||||||||
| Pump capacity flow rate | 22 ... 26 l/min | ||||||||||||||
| Pump capacity flow pressure | 0.4 ... 0.7 bar | ||||||||||||||
| Maximum suction | 0.2 ... 0.4 bar | ||||||||||||||
| Viscosity max. | 70 cST | ||||||||||||||
Refrigerant stage 1 |
Refrigerant | R452A | |||||||||||||
| Global Warming Potential for R452A | 2140 | ||||||||||||||
| Filling volume | 850 g | ||||||||||||||
| Carbon dioxide equivalent | 1.819 t | ||||||||||||||
| Heating capacity | 3 kW | ||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Cooling capacity (Medium: Ethanol) |
|
|
|||||||||||||
|
Cooling capacity
(Medium: Ethanol) |
|||||||||||||||
|
|||||||||||||||
| Pump capacity flow rate | 22 ... 26 l/min | ||||||||||||||
| Pump capacity flow pressure | 0.4 ... 0.7 bar | ||||||||||||||
| Maximum suction | 0.2 ... 0.4 bar | ||||||||||||||
| Viscosity max. | 70 cST | ||||||||||||||
Refrigerant stage 1 |
Refrigerant | R452A | |||||||||||||
| Global Warming Potential for R452A | 2140 | ||||||||||||||
| Filling volume | 850 g | ||||||||||||||
| Carbon dioxide equivalent | 1.819 t | ||||||||||||||
| Bath cover | integrated |
|---|
| Cooling of compressor | 1-stage Water |
|---|---|
| Cooling water pressure max. | 6 bar |
| Cooling water difference pressure | 3.5 ... 6 bar |
| Cooling water temperature max. | 20 °C |
| Cooling water consumption | 6.6 l/min |
| Classification according to DIN12876-1 |
Classification III (FL) |
|---|---|
| IP Code according to IEC 60529 |
IP 21 |
| Digital interface | Profibus optional |
|---|---|
| External pt100 sensor connection | integrated |
| Integrated programmer | 6x60 steps |
| Temperature control | ICC |
| Absolute temperature calibration | 3 Point Calibration |
| Temperature display | VFD |
| Temperature setting | Keypad |
| 400V/3PNPE/50Hz Plug 32A CEE | 9352756N.07 |
|---|---|
| 208V/3PPE/60Hz Without Plug | 9352756N.16 |
| 2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male) |
| 2 each barbed fittings for tubing 8 and 12 mm inner dia. (pump connections M16x1 female). |
Overview
- Dimensions and volumes
- Power and capacities
- Bath
- Cooling
- Other
- Electronics
- Available voltage versions
Accessory for FPW55-SL
Benefits
-
JULABO. Quality.
Highest standards of quality for a long product life. -
Green technology.
Development consistently applied environmentally friendly materials and technologies. -
Satisfied customers.
11 subsidiaries and more than 100 partners worldwide guarantee fast and qualified JULABO support. -
100% Checked.
100% testing. 100% quality. Each JULABO Circulator undergoes thorough quality testing before leaving the factory.
-
Quick start.
Individual JULABO consultation and comprehensive manuals at your disposal. -
Services 24/7.
Around the clock availability. You can find suitable accessories, data sheets, manuals, case studies, and more at www.julabo.com. -
Intelligent temperature control.
Intelligent cascade control - automatic and self-optimizing adaptation of the PID control parameters with external stability of +/- 0.05 °C. -
Connection of additional equipment
Stakei connections for solenoid valve, HSP booster pump and HST booster heater -
Early warning system for high/low temperature limits
Maximum safety for applications, optical and audible alarm, convertible to automated cut-off function -
Clever pump system
Reliable and consistent pump capacity, electronically adjustable pump stages -
Control from the external application
External Pt100 sensor connection for precise measurement and control directly in the external application -
For flammable bath fluid
Classification III (FL) according to DIN 12876-1 -
ATC3. Calibration.
‘Absolute Temperature Calibration’ for compensating a physically caused temperature difference, 3-point calibration. -
Process. Under control.
Full regulation of the dynamics control, access to all important control parameters for individual process optimization. -
100 % Cooling capacity
‘Active Cooling Control’ for cooling available throughout the entire working temperature range, fast cool-down even at higher temperatures -
Energy saving cooling
Proportional cooling control for automatic adjustment of cooling power or temporary switch-off of compressor as needed to save up to 90 % energy in comparison to unregulated cooling machines -
Condensation and ice protection
A heated cover plate prevents condensation or ice build-up in the bath