FPW52-SL Ultra-Low Refrigerated-Heating Circulator

for working temperatures from -95 °C to +150 °C with filling port upgradeable with additional heating and pump capacity for external control only

Ultra-Low Refrigerated Circulators with SL circulator impress with their high heating, cooling and pump capacities for external temperature control applications.

Request a quote
Data based on voltage version:
400V/3PNPE/50Hz (Plug 32A CEE)
FPW52-SL
Order No. 9 352 753N150.07
Working temperature range -60 ... +150 °C
Temperature stability ±0.05 °C
Heating capacity 3 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -40 -60
3 2.8 1.6 0.65 0.1
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -40 -60
kW 3 2.8 1.6 0.65 0.1
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Filling volume 24 l
Dimensions
(W × L × H)
59 x 76 x 116 cm
FPW52-SL
Order No. 9 352 753N150.16
Working temperature range -60 ... +150 °C
Temperature stability ±0.05 °C
Heating capacity 3 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -40 -60
3 2.8 1.6 0.65 0.1
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -40 -60
kW 3 2.8 1.6 0.65 0.1
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Filling volume 24 l
Dimensions
(W × L × H)
59 x 76 x 116 cm

Product Features

  • VFD COMFORT DISPLAY
  • LCD DIALOG DISPLAY backlit for convenient interactive operation
  • Keypad for setpoints, warning/safety values and menu functions
  • ICC (Intelligent Cascade Control), self-optimizing temperature control
  • TCF Temperature Control Features to optimize the control behavior
  • ATC3 3-Point-Calibration
  • Pt100 External sensor connection for measurement and control
  • SMART PUMP, electronically adjustable pump stages
  • Adjustable high temperature cut-out, visible via display
  • RS232/RS485 interface for online communication
  • Integrated programmer for 6 x 60 program steps
  • Connections for solenoid valve and HSP booster pump
  • Proportional cooling control
Dimensions and volumes
Dimensions
(W × L × H)
59 x 76 x 116 cm
Filling volume 24 l
Weight 153 kg
Pump connections M16x1  male
Cooling Water Connection G¾ in
Temperature values
Working temperature range -60 ... +150 °C
Temperature stability ±0.05 °C
Setting the resolution of the temperature display 0.01 °C
Temperature display resolution 0.01 °C
Ambient temperature +5.0 ... +40.0 °C
Data based on voltage version:
400V/3PNPE/50Hz (Plug 32A CEE)
Performance values
Heating capacity 3 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -40 -60
3 2.8 1.6 0.65 0.1
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -40 -60
kW 3 2.8 1.6 0.65 0.1
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Maximum suction 0.2 ... 0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R452A
Global Warming Potential for R452A 2140
Filling volume 850 g
Carbon dioxide equivalent 1.819 t
Performance values
Heating capacity 3 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -40 -60
3 2.8 1.6 0.65 0.1
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -40 -60
kW 3 2.8 1.6 0.65 0.1
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Maximum suction 0.2 ... 0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R452A
Global Warming Potential for R452A 2140
Filling volume 850 g
Carbon dioxide equivalent 1.819 t
Bath
Bath cover integrated
Cooling
Cooling of compressor 1-stage Water
Cooling water pressure max. 6 bar
Cooling water difference pressure 3.5 ... 6 bar
Cooling water temperature max. 20 °C
Cooling water consumption 5 l/min
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Electronics
Digital interface Profibus optional
External pt100 sensor connection integrated
Integrated programmer 6x60 steps
Temperature control ICC
Absolute temperature calibration 3 Point Calibration
Temperature display VFD
Temperature setting Keypad
Available voltage versions
400V/3PNPE/50Hz Plug 32A CEE 9352753N150.07
208V/3PPE/60Hz Without Plug 9352753N150.16
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)
2 each barbed fittings for tubing 8 and 12 mm inner dia. (pump connections M16x1 female).

Accessory for FPW52-SL

Brochure

TopTech - HighTech Circulators
English
PDF - 3.93 MB

Download

Operation Manual

Betriebsanleitung + Bedienungsanleitung
English
PDF - 9.69 MB

Download

Product Data Sheet

FPW52-SL
English
PDF - 334.6 KB

Download

We would also be happy to send you our printed catalogs and brochures by post.

Request brochure

Benefits

  • Made in Germany
    JULABO. Quality.
    Highest standards of quality for a long product life.
  • Blatt wächst in einer Glühbirne
    Green technology.
    Development consistently applied environmentally friendly materials and technologies.
  • network
    Satisfied customers.
    11 subsidiaries and more than 100 partners worldwide guarantee fast and qualified JULABO support.
  • 100% geprüft
    100% Checked.
    100% testing. 100% quality. Each JULABO Circulator undergoes thorough quality testing before leaving the factory.
  • Quickstart
    Quick start.
    Individual JULABO consultation and comprehensive manuals at your disposal.
  • Downloads
    Services 24/7.
    Around the clock availability. You can find suitable accessories, data sheets, manuals, case studies, and more at www.julabo.com.
  • ICC
    Intelligent temperature control.
    Intelligent cascade control - automatic and self-optimizing adaptation of the PID control parameters with external stability of +/- 0.05 °C.
  • TCF
    Connection of additional equipment
    Stakei connections for solenoid valve, HSP booster pump and HST booster heater
  • EarlyWarningSystemLowLiquid.ai
    Early warning system for high/low temperature limits
    Maximum safety for applications, optical and audible alarm, convertible to automated cut-off function
  • Intelligentes Pumpensystem
    Clever pump system
    Reliable and consistent pump capacity, electronically adjustable pump stages
  • PT100
    Control from the external application
    External Pt100 sensor connection for precise measurement and control directly in the external application
  • S3
    For flammable bath fluid
    Classification III (FL) according to DIN 12876-1
  • ATC3.eps
    ATC3. Calibration.
    ‘Absolute Temperature Calibration’ for compensating a physically caused temperature difference, 3-point calibration.
  • TCF.eps
    Process. Under control.
    Full regulation of the dynamics control, access to all important control parameters for individual process optimization.
  • ACC
    100 % Cooling capacity
    ‘Active Cooling Control’ for cooling available throughout the entire working temperature range, fast cool-down even at higher temperatures
  • TCF
    Energy saving cooling
    Proportional cooling control for automatic adjustment of cooling power or temporary switch-off of compressor as needed to save up to 90 % energy in comparison to unregulated cooling machines
  • TCF
    Condensation and ice protection
    A heated cover plate prevents condensation or ice build-up in the bath