FPW52-SL Ultra-Low Refrigerated / Heating Circulator
with filling port upgradeable with additional heating and pump capacity for external control only
Performance data depends on the mains voltage in the country of use. Please select:
Country of use | Mains voltage | Plug | JULABO voltage version code
Order No. | 9352753N150.07 | ||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -55 ... +150 °C | ||||||||||||||||||||||
Temperature stability | ±0.05 °C | ||||||||||||||||||||||
Heating capacity | 3 kW | ||||||||||||||||||||||
Cooling capacity
* Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
|
|||||||||||||||||||||||
Pump capacity flow rate | 22 ... 26 l/min | ||||||||||||||||||||||
Pump capacity flow pressure | 0.4 ... 0.7 bar | ||||||||||||||||||||||
Filling volume | 17 ... 24 l | ||||||||||||||||||||||
Dimensions (W × L × H) |
59 x 76 x 116 cm |
Product Features
- VFD COMFORT DISPLAY
- LCD DIALOG DISPLAY backlit for convenient interactive operation
- Keypad for setpoints, warning/safety values and menu functions
- ICC (Intelligent Cascade Control), self-optimizing temperature control
- TCF Temperature Control Features to optimize the control behavior
- ATC3 3-Point-Calibration
- Pt100 External sensor connection for measurement and control
- SMART PUMP, electronically adjustable pump stages
- Unique early warning system for low liquid level
- Adjustable high temperature cut-out, visible via display
- RS232/RS485 interface for online communication
- Integrated programmer for 6 x 60 program steps
- Connections for solenoid valve and HSP booster pump
- Proportional cooling control
- Active Cooling Control
Dimensions (W × L × H) |
59 x 76 x 116 cm |
---|---|
Filling volume | 17 ... 24 l |
Weight | 153 kg |
Pump connections | M16x1 male |
Cooling Water Connection | G¾ in |
Working temperature range | -55 ... +150 °C |
---|---|
Temperature stability | ±0.05 °C |
Setting the resolution of the temperature display | 0.01 °C |
Temperature display resolution | 0.01 °C |
Ambient temperature | +5 ... +40 °C |
Performance data depends on the mains voltage in the country of use. Please select:
Country of use | Mains voltage | Plug | JULABO voltage version code
Heating capacity | 3 kW | ||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity
* Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
|
|||||||||||||||||||||||
Pump capacity flow rate | 22 ... 26 l/min | ||||||||||||||||||||||
Pump capacity flow pressure | 0.4 ... 0.7 bar | ||||||||||||||||||||||
Maximum suction | 0.2 ... 0.4 bar | ||||||||||||||||||||||
Viscosity max. | 70 cSt | ||||||||||||||||||||||
Power consumption | 20 A |
Bath cover | integrated |
---|
Cooling of compressor | 1-stage Water |
---|---|
Cooling water pressure max. | 6 bar |
to provide full cooling capacity according to catalog values | |
Cooling water temperature range | 20 °C |
---|---|
Cooling water difference pressure | 3.5 bar |
Cooling water consumption | 5 l/min * |
* Cooling water consumption may vary outside recommended cooling water properties. |
Classification according to DIN12876-1 |
Classification III (FL) |
---|---|
IP Code according to IEC 60529 |
IP 21 |
Interfaces | Profibus optional |
---|---|
External pt100 sensor connection | integrated |
Integrated programmer | 6x60 steps |
Temperature control | ICC |
Absolute temperature calibration | 3 Point Calibration |
Temperature display | VFD |
Temperature setting | Keypad |
9352753N150.07 | 400V/3PNPE/50Hz (Plug 32A CEE) (R452A) |
---|---|
9352753N150.16 | 208V/3PPE/60Hz (Without Plug) (R452A) |
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male) |
Accessory for FPW52-SL
Benefits
-
For flammable bath fluid
Classification III (FL) according to DIN 12876-1 -
ATC3. Calibration.
‘Absolute Temperature Calibration’ for compensating a physically caused temperature difference, 3-point calibration. -
Process. Under control.
Full regulation of the dynamics control, access to all important control parameters for individual process optimization. -
100 % Cooling capacity
‘Active Cooling Control’ for cooling available throughout the entire working temperature range, fast cool-down even at higher temperatures