Print

PRESTO A45 New

Temperature Control System / Process System

Reactor temperature control, tests for all kinds of substances or temperature simulation – the new PRESTO are made for highly precise temperature control and rapid temperature changes. Highly efficient components allow extremely fast  compensation of exothermic and endothermic reactions.  Lab users benefit from high flow rates, constant pressure, and a controlled build-up of pump pressure. Permanent internal monitoring and self-lubricating pumps contribute to the new PRESTO’s long service life. The integrated 5.7’’ industrial touch screen displays all important information clearly and concisely enhancing ease of use considerably.

Your advantages

  • For highly precise, external temperature applications
  • Rapid heating and cooling
  • Wide working temperature ranges without changing fluids
  • Highest performance with small footprint
  • Space-saving design optimizes space utilization in your lab
  • 5,7” industrial color TFT touch screen
  • USB (Host und Device)
  • Ethernet
  • SD-Card slot
  • RS232
  • Modbus
  • RS485 (Accessory)
  • Profibus DP (Accessory)
  • Stand-by input
  • Removable venting grid for simplified removal of dust

Technical Data

Order No. 9420452
Category Temperature Control PRESTO
Working temperature range (°C) -45 ... +250
Temperature control ICC
Temperature stability (°C) ±0.05 ... ±0.1
Setting / display resolution 0.01 °C
Integrated programmer 8x60 steps
Temperature Display TFT Touchscreen
Heating capacity (kW) 6 @ 400V; 6 @ 230V; 5.5 @ 208V
Cooling capacity (Medium: JULABO Thermal | Ethanol)
°C200200-10-20-30-40
kW3.43.53.32.61.810.3
Pump capacity flow rate (l/min) 35 ... 76
Pump capacity flow pressure (bar) 0.48 ... 3.2 @ 400V; 0.48 ... 3.2 @ 230V; 0.48 ... 3 @ 208V
Viscosity max. (cSt) 50
Pump connections M30x1.5
Refrigerant stage 1 R452A
Filling volume refrigerant stage 1 (g) 1320
Global Warming Potential for R452A 2140
Carbon dioxide equivalent stage 1 (t) 2.825
External Pt100 sensor connection integrated
Digital interfaces RS232
SD memory card
USB
Ethernet
Modbus
Alarm-out
Optional: RS485, Profibus
Analog connection input / output Optional
Ambient temperature 5 ... 40 °C
Dimensions W x L x H (cm) 53 x 66.5 x 126
Weight (kg) 210
Sound pressure level (distance 1 m) max. (dBA) 69
Process volume min. (active heat exchanger volume) liters 7.5 (3.5)
Internal usable expansion vol. (liters) 7.5
Classification according to DIN12876-1 Classification III (FL)
Cooling of compressor 1-stage Air
Available voltage versions 3 x 208-230V/60Hz (+/- 5%) / 23A @ 208V, 25A @ 230V / Without Plug
3 x 230V/50Hz (+/- 10%) / 25A / Plug 32A CEE
3 x 400V/50Hz (+/- 10%) / 13A / Plug 16A CEE

Characteristics

Display 
State-of-the-art display technology
State-of-the-art display technology
TFT Display for comfortable user guidance, colored display of measurement values, graphs and control options, user-defined views
Operation 
Optimal ease of use
Optimal ease of use
Touch screen for direct operation via display
Instructions inside
Instructions inside
Help menus and explanations in plain text for all control options, help messages and warning messages
Multilingual user guidance
Multilingual user guidance
Language selection for display of control options, notifications and warning messages via touchscreen
Convenience for several users
Convenience for several users
Administrator level for customizing instrument settings, user levels with limited permissions for fast and safe defined access, password protection, all levels adjustable
Temperature Control
For perfect results
For perfect results
Intelligent Cascade Control’, automatic & self optimizing adjustment of PID control parameters, temperature stability ±0.01 °C ... <±0.2 °C
Full control
Full control
Temperature Control Features’, for individual optimization, access to all important control parameters, additional settings for band limit, limits, co-speedfactor etc.
Control from the external application
Control from the external application
External Pt100 sensor connection for precise measurement and control directly in the external application
Highest measuring accuracy
Highest measuring accuracy
Absolute Temperature Calibration’ for manual compensation of a temperature difference, 3-point calibration
Refrigeration Technology
Consistent cooling capacity
Consistent cooling capacity
Easily removable venting grid for quick and easy cleaning
100 % Cooling capacity
100 % Cooling capacity
Active Cooling Control’ for cooling available throughout the entire working temperature range, fast cool-down even at higher temperatures
Energy saving cooling
Energy saving cooling
Proportional cooling control for automatic adjustment of cooling power or temporary switch-off of compressor as needed to save up to 90 % energy in comparison to unregulated cooling machines
Technical Features
Intelligent pump system
Intelligent pump system
Reliable and consistent pump capacity, electronically adjustable pump stages or pressure value, automatic adjustment of pump capacity to viscosity
Communication via networks
Communication via networks
For the remote control of instruments via Ethernet networks, full access to all functions of the unit via a networkcapable PC
Intelligent communication
Intelligent communication
USB connection for data exchange (e.g. service data) or for wireless remote control via WirelessTEMP®
Data exchange via SD-Card
Data exchange via SD-Card
For data exchange (e.g. service data) via SD memory card
Serial connection
Serial connection
RS232 interface for PC connection, e.g. for data communication and recording of measured values
Comfortable program control
Comfortable program control
Integrated programmer for the execution of time and temperature dependant profiles, 8 temperature profiles with 60 steps max., with real time clock
Quiet as a whisper
Quiet as a whisper
Efficient components produce only a minimal sound decibel level
Space-saving footprint
Space-saving footprint
All connections as well supply and exhaust air are located at the front or rear, no venting grids on the sides, units can be placed close to each other or the application
Continuous operation up to +40 °C
Continuous operation up to +40 °C
Robust temperature control instrument, continuous operation even at ambient temperatures of up to +40 °C
Easy transport by one person
Easy transport by one person
Ergonomic design facilitates moving and positioning by one person
Filling level at a glance
Filling level at a glance
Backlit indicator for selected pump stages and filling volume
Warning & Safety Functions
Early warning system for high/low temperature limits
Early warning system for high/low temperature limits
Maximum safety for applications, optical and audible signal when limits are exceeded.
Duplicate safety
Duplicate safety
Adjustable high temperature cut-off for internal tank and for integrated expansion vessel
For flammable bath fluid
For flammable bath fluid
Classification III (FL) according to DIN 12876-1
Quick support
Quick support
If an error occurs, the integrated Black-Box function permits fast diagnosis by the JULABO service team

Technical changes without prior notification.
Images may deviate from the original.

We would be pleased to submit a non-binding offer for the selected product or answer all pertinent questions.

click here for quotation request PRESTO A45
 

Contact

JULABO GmbH
Gerhard-Juchheim-Strasse 1
77960 Seelbach / Germany
Tel: +49 7823 51-0
info.de@julabo.com
www.julabo.com
Contact form

Catalog request
Order our brochures!