Print

HT60-M3-CU High Temperature Circulator

2-zone technique for working temperatures up to 400 °C

JULABO high temperature circulators: The response to present requirements in the market. Peak performance, rapid heating and cooling, fast action in case of exotherm reactions - these are only few of the features of our high temperature circulators. The self-optimizing controller (ICC - Intelligent Cascade Control) ensures high temperature stability. The classic PID control parameters are available at the touch of a button for modification at any time. MULTI-DISPLAY (LED) and interactive LCD display offer convenient operation. The implemented heating automation, strong pump and controlled cooling produce rapid and precise results for temperature applications to +4 00 °C. The instruments are completely sealed. This prevents the release of oil fumes and results in long life time of the heat transfer liquid. Automatic air-purge allows easy filling. Electrical connections and the optional Profibus permit future-oriented automation technology. JULABO high temperature circulators are recommended for applications using jacketed reactors and autoclaves, distillation columns as well as laboratory, miniplant and process-scale applications.  

Your advantages

  • Working temperature range up to 400 °C
  • Self-optimizing ICC temperature controller
  • Strong pumping capacity 18 l/min - 1.2 bar max
  • Automatic air-purge allows easy filling
  • Automated heat-up feature adjusts to external systems
  • Bright MULTI-DISPLAY (LED) with display resolution 0.01 °C
  • Backlit interactive 4-line LCD display
  • On-line communication via built-in RS232 interface
  • Stand-by input for external emergency switch-off
  • Integrated programmer stores up to 6 profiles (60 sections each)
  • Electronic level indication
  • ATC function for simple correction of temperature variations
  • ATC function allows use as testing instrument according to ISO 9000
  • Tap water cooling for temperatures from 40 to 400 °C and rapid cooling
  • Cooling capacity up to 15 kW (with tap water 20°C)
     

Technical Data

Order No. 9800066
Category High Temperature Circulators Forte HT
Working temperature range (°C) +40 ... +400
Temperature stability (°C) ±0.01-0.1
Setting / display resolution 0.01 °C
Integrated programmer 6x60 steps
Temperature Display LCD , LED
Heating capacity (kW) 6
Cooling capacity
°C20
kW15
Pump capacity flow rate (l/min) 14-18
Pump capacity flow pressure (bar) 0.8-1.2
Pump connections M16x1
Filling volume (liters) 2
External Pt100 sensor connection integrated
Digital interfaces RS232
RS485
Optional Profibus
Ambient temperature 5...40 °C
Dimensions W x L x H (cm) 43 x 23 x 58
Weight (kg) 35
Classification according to DIN12876-1 Classification III (FL)
Display for pump pressure and filling volume Backlit indicator
Filling volume expansion vessel liters 1.6+0.9
Dimensions Control unit B x T x H 25 x 25 x 18

Characteristics

Display 
Easy to read
Easy to read
Large LED temperature display for actual value and setpoint (resolution 0.1 °C)
Additional plain text information
Additional plain text information
Comfortable LCD dialog display for interactive operation with plain text display
Pump stage and liquid level
Pump stage and liquid level
Backlit indicator for selected pump stages and filling volume on Presto® PLUS, Magnum 91 & Forte HT
Operation 
Clearly structured
Clearly structured
Comfortable, splash-proof keypad for setpoint adjustment, high/low temperatures, timer and shaking frequeny (SW models)
Temperature Control
For perfect results
For perfect results
Intelligent Cascade Control’, automatic & self optimizing adjustment of PID control parameters, temperature stability ±0.005 °C internal, <±0.05 °C external
Full control
Full control
Temperature Control Features’ for individual optimization, access to all important control parameters, additional settings for band limit, limits, Co-Speedfactor etc.
High measuring accuracy
High measuring accuracy
Absolute Temperature Calibration’ for manual compensation of a temperature difference, 1-point calibration
Technical Features
Control from the external application
Control from the external application
External Pt100 sensor connection for precise measurement and control directly in the external application
Connection compliant to standard
Connection compliant to standard
RS232/RS485 dual-interface for serial data transmission according to EIA-485 industry standard (2-wire bus technology), upgradable with Profibus DP
Optimal program control
Optimal program control
For the execution of time and temperature dependant profiles, 6 temperature profiles with 60 steps max., with real time clock
Warning & Safety Functions
Early warning system for high/low temperature limits
Early warning system for high/low temperature limits
Maximum safety for applications, optical and audible alarm, convertible to automated cut-off function
Enhanced protective functions 
Enhanced protective functions 
Maximum safety, adjustable high temperature cut-off or dry-running protection, additional display of setpoints permits easy and precise adjustments
For flammable bath fluids
For flammable bath fluids
Classification III (FL) according to DIN 12876-1

Other characteristics

Permissible cooling water pressure 4.5 bar max, pressure difference between IN and OUT: 2 to 4.5 bar

Technical changes without prior notification.
Images may deviate from the original.

We would be pleased to submit a non-binding offer for the selected product or answer all pertinent questions.

click here for quotation request HT60-M3-CU
 

Contact

JULABO GmbH
Gerhard-Juchheim-Strasse 1
77960 Seelbach / Germany
Tel: +49 7823 51-0
info.de@julabo.com
www.julabo.com
Contact form

Available voltage versions

Model: HT60-M3-CU
Order No.: 9800066
208 V / 3 Ph. / 60 Hz
230 V / 3 Ph. / 60 Hz
Catalog request
Order our brochures!