Special solution - temperature cycling test on gas springs
08/13/2024
JULABO solution:
Two stainless steel double-jacketed vessels were manufactured. The jacket area is temperature-controlled with silicon oil and the internal temperature is measured using an external sensor. A PRESTO W91 is used on the first vessel for cooling down to -40°C and an SL12 circulator with booster heater is used on the second vessel for heating up to +100°C.
The gas springs are now alternately immersed in the temperature-controlled vessels. To ensure better heat transfer, the suspension of the springs is constantly rotated. This creates a circular flow in the vessel.
The vessels are also installed in a vapor-tight housing to reduce dilution of the medium.