Case Study: PRESTO W50t Process system
Heating up of an QVF 30 l reactor from +20 °C to +150 °C with a PRESTO W50t Process system
Objective
This case study tests the heating capacity of the JULABO PRESTO W50t Process system with a 30 l glass reactor. The PRESTO W50t is connected to the reactor via two 2 m metal hoses. The PRESTO W50t Process system has been programmed to heat up from +20 °C to +150 °C.
Test Conditions
| Product | PRESTO W50t Process system | ||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||||
| Heating capacity | 12 kW | ||||||||||||||||
| Pump pressure | 0.5 bar | ||||||||||||||||
| Duration | 49 hr:min | ||||||||||||||||
| Test volume | 22.5 l | ||||||||||||||||
| Setpoint 1 | +20 °C | ||||||||||||||||
| Setpoint 2 | +150 °C | ||||||||||||||||
| Reactor | 30l QVF (glass) | ||||||||||||||||
| Tubing length | 2 m | ||||||||||||||||
| Fluid in unit | Thermal HL60 | ||||||||||||||||
| Fluid in reactor | Thermal H20S | ||||||||||||||||
Environment
| Room temperature | 20 °C |
|---|---|
| Humidity | 45 % |
| Voltage version | 400V/3PNPE/50Hz |
Control Parameters
| XP | 0.2 K |
| TN | 695 s |
| TV | 92 s |
| XPU | 50.88 K |
Test Results
The PRESTO W50t Process system heats up the reactor from +20 °C to +150 °C in 49 min.
| Set point |
| Temperature in the application |
| Temperature in temperature control unit |