Case Study: MAGIO MS-1200F Refrigerated / heating circulator

Cooling a Normag 2 l reactor from +150 °C to lowest possible temperature with a MAGIO MS-1200F Refrigerated / heating circulator

Objective

This case study tests the lowest temperature of the JULABO MAGIO MS-1200F Refrigerated / heating circulator with a 2 l glass reactor. The MAGIO MS-1200F is connected to the reactor via two 1 m metal hoses. The MAGIO MS-1200F Refrigerated / heating circulator cools from +150 °C to -34 °C maximum low temperature.

Test Conditions

Product MAGIO MS-1200F Refrigerated / heating circulator
Cooling capacity (Ethanol)
°C 20 0 -20 -30 -40
kW 1.26 1.11 0.64 0.39 0.22
Heating capacity 2 kW
Pump pressure 0.5 bar Band limit °k
Duration 2:55 hr:min Test volume 2 l
Setpoint 1 +150 °C Setpoint 2 -50 °C
Reactor 2l Normag (glass) Tubing length 1 m
Fluid in unit Thermal H10 Fluid in reactor Thermal H10

Environment

Room temperature 22 °C
Humidity 45 %
Voltage version 230V/50Hz

Control Parameters

XP 0.2 K
TN 450 s
TV 15 s
XPU 2.5 K
MAGIO MS-1200F details

Test Results

The MAGIO MS-1200F Refrigerated / heating circulator cools the reactor from +150 °C to the lowest temperature of -34 °C in 2:55 h.

  • Set point
  • Temperature in the application
  • Temperature in temperature control unit