Case Study: PRESTO A85t Process system
Cooling a Büchiglas 50 l reactor from +20 °C to lowest possible temperature with a PRESTO A85t Process system
Objective
This case study tests the lowest temperature of the JULABO PRESTO A85t Process system with a 50 l glass reactor. The PRESTO A85t is connected to the reactor via two 2 m metal hoses. The PRESTO A85t Process system cools from +20 °C to -73 °C maximum low temperature.
Test Conditions
| Product | PRESTO A85t Process system | ||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||||||
| Heating capacity | 15 kW | ||||||||||||||||||
| Pump pressure | 0.5 bar | Band limit | °k | ||||||||||||||||
| Duration | 6:30 hr:min | Test volume | 50 l | ||||||||||||||||
| Setpoint 1 | +20 °C | Setpoint 2 | -85 °C | ||||||||||||||||
| Reactor | 50l Büchiglas (glass) | Tubing length | 2 m | ||||||||||||||||
| Fluid in unit | Thermal HL80 | Fluid in reactor | Thermal HL80 | ||||||||||||||||
Environment
| Room temperature | 20 °C |
|---|---|
| Humidity | 45 % |
| Voltage version | 400V/3PNPE/50Hz |
Control Parameters
| XP | 0.2 K |
| TN | 720 s |
| TV | 100 s |
| XPU | 24 K |