F81-HL Ultra-Low Refrigerated / Heating Circulator

low temperatures below -50 °C with bath opening for internal and external applications

The Ultra-Low Refrigerated Circulators of the HighTech Series with HL or SL circulator feature powerful pressure and suction pumps. The in-struments provide the entire variety of functions of the professional HighTech Series of circulators.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
F81-HL
Order No. 9312681.33
Working temperature range -81 ... +100 °C
Temperature stability ±0.02 °C
Heating capacity 1.3 kW
Cooling capacity
Medium: Ethanol
°C 20 0 -20 -40 -60 -80
kW* 0.45
0.38
0.36
0.32
0.27
0.07
Refrigerant Information Refrigerant stage 1 Refrigerant stage 2
Refrigerant R404A R23
Global Warming Potential 3922 14800
Filling weight 500 g 120 g
Carbon dioxide equivalent 1.961 t 1.776 t
* Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Usable bath opening
(W x L / D)
13 x 15 / 16 cm
Filling volume 5 ... 6.5 l
Dimensions
(W × L × H)
50 x 58 x 89 cm
Arrange callback Accessory

Product Features

  • LCD DIALOG DISPLAY backlit for convenient interactive operation
  • ICC (Intelligent Cascade Control), self-optimizing temperature control
  • TCF Temperature Control Features to optimize the control behavior
  • RS232/RS485 interface for online communication
  • Integrated programmer for 6 x 60 program steps
  • Connections for solenoid valve and HSP booster pump
  • Proportional cooling control
Dimensions and volumes
Dimensions
(W × L × H)
50 x 58 x 89 cm
Filling volume 5 ... 6.5 l
Weight 88 kg
Pump connections M16x1 male
Temperature values
Working temperature range -81 ... +100 °C
Temperature stability ±0.02 °C
Setting the resolution of the temperature display 0.01 °C
Temperature display resolution 0.01 °C
Ambient temperature +5 ... +40 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 1.3 kW
Cooling capacity
Medium: Ethanol
°C 20 0 -20 -40 -60 -80
kW* 0.45
0.38
0.36
0.32
0.27
0.07
Refrigerant Information Refrigerant stage 1 Refrigerant stage 2
Refrigerant R404A R23
Global Warming Potential 3922 14800
Filling weight 500 g 120 g
Carbon dioxide equivalent 1.961 t 1.776 t
* Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Maximum suction 0.2 ... 0.4 bar
Viscosity max. 70 cST
Power 16 A
Bath
Bath cover integrated
Usable bath opening
(W x L / D)
13 x 15 / 16 cm
Cooling
Cooling of compressor 2-stage Air
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Electronics
Interfaces Profibus optional
External pt100 sensor connection integrated
Integrated programmer 6x60 steps
Temperature control ICC
Absolute temperature calibration 3 Point Calibration
Temperature display VFD
Temperature setting Keypad
Available voltage versions
9312681.33 230V/50-60Hz (Schuko Plug - CEE 7/4 Plug Type F) (R404A)
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)

Accessory for F81-HL

Contact expert

We would also be happy to send you our printed catalogs and brochures by post.


Request brochure

Benefits

  • 100% geprüft
    100% Checked.
    100% testing. 100% quality. Each JULABO Circulator undergoes thorough quality testing before leaving the factory.
  • Blatt wächst in einer Glühbirne
    Green technology.
    Development consistently applied environmentally friendly materials and technologies.
  • Made in Germany
    JULABO. Quality.
    Highest standards of quality for a long product life.
  • Quickstart
    Quick start.
    Individual JULABO consultation and comprehensive manuals at your disposal.