FPW50-HE Refrigerated - Heating Circulator

for working temperatures to +200°C

The HighTech Series features refrigerated heating /circulators with innovative technology for sophisticated applications. The instruments provide powerful, electronically adjustable pressure and suction pumps.

The instruments can be used for internal as well as external temperature control applications of closed and open systems.

Request a quote
Data based on voltage version:
230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F)
FPW50-HE
Order No. 9 212 651.13
Working temperature range -50 ... +200 °C
Temperature stability ±0.01 °C
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -30 -40
0.9 0.8 0.4 0.16 0.03
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -30 -40
kW 0.9 0.8 0.4 0.16 0.03
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Usable bath opening
(W x L / D)
18 x 12 / 15 cm
Filling volume 8 l
Dimensions
(W × L × H)
42 x 49 x 72 cm
FPW50-HE
Order No. 9 212 651.03
Working temperature range -50 ... +200 °C
Temperature stability ±0.01 °C
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -30 -40
0.9 0.8 0.49 0.26 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -30 -40
kW 0.9 0.8 0.49 0.26 0.11
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Usable bath opening
(W x L / D)
18 x 12 / 15 cm
Filling volume 8 l
Dimensions
(W × L × H)
42 x 49 x 72 cm

Product Features

  • VFD COMFORT DISPLAY
  • Keypad for setpoints, warning/safety values and menu functions
  • ICC (Intelligent Cascade Control), self-optimizing temperature control
  • ATC3 3-Point-Calibration
  • Pt100 External sensor connection for measurement and control
  • SMART PUMP, electronically adjustable pump stages
  • Adjustable pressure and suction pump
  • Adjustable high temperature cut-out, visible via display
  • RS232 interface for online communication
  • Optional: analogue interfaces
  • Integrated programmer for 10 program steps
  • Active Cooling Control
Dimensions and volumes
Dimensions
(W × L × H)
42 x 49 x 72 cm
Filling volume 8 l
Weight 57 kg
Pump connections M16x1  male
Cooling Water Connection G¾ in
Barbed fittings inner diameter 8/12 mm
Temperature values
Working temperature range -50 ... +200 °C
Temperature stability ±0.01 °C
Temperature stability external ±0.01 ... ±0.2 °C
Setting the resolution of the temperature display 0.01 °C
Temperature display resolution 0.01 °C
Ambient temperature +5.0 ... +40.0 °C
Data based on voltage version:
230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F)
Performance values
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -30 -40
0.9 0.8 0.4 0.16 0.03
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -30 -40
kW 0.9 0.8 0.4 0.16 0.03
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Maximum suction 0.2 ... 0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R449A
Global Warming Potential for R449A 1397
Filling volume 480 g
Carbon dioxide equivalent 0.671 t
Performance values
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20 -30 -40
0.9 0.8 0.49 0.26 0.11
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20 -30 -40
kW 0.9 0.8 0.49 0.26 0.11
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 0.4 ... 0.7 bar
Maximum suction 0.2 ... 0.4 bar
Viscosity max. 70 cST
Refrigerant stage 1
Refrigerant R452A
Global Warming Potential for R452A 2140
Filling volume 480 g
Carbon dioxide equivalent 1.027 t
Bath
Bath cover integrated
Usable bath opening
(W x L / D)
18 x 12 / 15 cm
Cooling
Cooling of compressor 1-stage Water
Cooling water pressure max. 6 ... 6 bar
Cooling water difference pressure 3.5 ... 6 bar
Cooling water temperature max. 20 °C
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump type Immersion Pump
Electronics
Digital interface Profibus optional
External pt100 sensor connection integrated
Integrated programmer 1x10 steps
Temperature control ICC
Absolute temperature calibration 3 Point Calibration
Temperature display VFD
Temperature setting Keypad
Available voltage versions
230V/60Hz Nema N6-20 Plug 9212651.13
230V/50Hz Schuko Plug - CEE 7/4 Plug Type F 9212651.03
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)

Accessory for FPW50-HE

Brochure

TopTech - HighTech Circulators
English
PDF - 3.93 MB

Download

Operation Manual

Betriebsanleitung + Bedienungsanleitung
English
PDF - 8.16 MB

Download

Product Data Sheet

FPW50-HE
English
PDF - 292.16 KB

Download

We would also be happy to send you our printed catalogs and brochures by post.

Request brochure

Benefits

  • Made in Germany
    JULABO. Quality.
    Highest standards of quality for a long product life.
  • Blatt wächst in einer Glühbirne
    Green technology.
    Development consistently applied environmentally friendly materials and technologies.
  • network
    Satisfied customers.
    11 subsidiaries and more than 100 partners worldwide guarantee fast and qualified JULABO support.
  • 100% geprüft
    100% Checked.
    100% testing. 100% quality. Each JULABO Circulator undergoes thorough quality testing before leaving the factory.
  • Quickstart
    Quick start.
    Individual JULABO consultation and comprehensive manuals at your disposal.
  • Downloads
    Services 24/7.
    Around the clock availability. You can find suitable accessories, data sheets, manuals, case studies, and more at www.julabo.com.
  • EarlyWarningSystemLowLiquid.ai
    Early warning system for high/low temperature limits
    Maximum safety for applications, optical and audible alarm, convertible to automated cut-off function
  • Intelligentes Pumpensystem
    Clever pump system
    Reliable and consistent pump capacity, electronically adjustable pump stages
  • PT100
    Control from the external application
    External Pt100 sensor connection for precise measurement and control directly in the external application
  • S3
    For flammable bath fluid
    Classification III (FL) according to DIN 12876-1
  • ATC
    ATC3. Calibration.
    ‘Absolute Temperature Calibration’ for compensating a physically caused temperature difference, 3-point calibration.
  • TCF
    Process. Under control.
    Full regulation of the dynamics control, access to all important control parameters for individual process optimization.
  • ACC
    100 % Cooling capacity
    ‘Active Cooling Control’ for cooling available throughout the entire working temperature range, fast cool-down even at higher temperatures
  • TCF
    Energy saving cooling
    Proportional cooling control for automatic adjustment of cooling power or temporary switch-off of compressor as needed to save up to 90 % energy in comparison to unregulated cooling machines