CF30 Cryo-Compact Circulator

The Cryo-Compact Circulators of the CF Series provide powerful cooling and heating capabilities in a space-saving, compact design. The instru-ments feature 2 kW heating capacity and classification III according to DIN 12876-1. The Cryo-Compact Circulators have pump connections for external temperature control applications and a bath opening for tem-perature control of small objects.

Request a quote
Data based on voltage version:
230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F)
CF30
Order No. 9 400 330.13
Working temperature range -30 ... +150 °C
Temperature stability ±0.03 °C
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20
0.32 0.25 0.15
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20
kW 0.32 0.25 0.15
Pump capacity flow rate 15 l/min
Pump capacity flow pressure 0.35 bar
Usable bath opening
(W x L / D)
16 x 3 / 14 cm
Filling volume 2 ... 3 l
Dimensions
(W × L × H)
24 x 46 x 40 cm
CF30
Order No. 9 400 330.02
Working temperature range -30 ... +150 °C
Temperature stability ±0.03 °C
Heating capacity 1 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20
0.32 0.25 0.15
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20
kW 0.32 0.25 0.15
Pump capacity flow rate 15 l/min
Pump capacity flow pressure 0.35 bar
Usable bath opening
(W x L / D)
16 x 3 / 14 cm
Filling volume 2 ... 3 l
Dimensions
(W × L × H)
24 x 46 x 40 cm
CF30
Order No. 9 400 330.03
Working temperature range -30 ... +150 °C
Temperature stability ±0.03 °C
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20
0.32 0.25 0.15
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20
kW 0.32 0.25 0.15
Pump capacity flow rate 15 l/min
Pump capacity flow pressure 0.35 bar
Usable bath opening
(W x L / D)
16 x 3 / 14 cm
Filling volume 2 ... 3 l
Dimensions
(W × L × H)
24 x 46 x 40 cm
CF30
Order No. 9 400 330.04
Working temperature range -30 ... +150 °C
Temperature stability ±0.03 °C
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20
0.32 0.25 0.15
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20
kW 0.32 0.25 0.15
Pump capacity flow rate 15 l/min
Pump capacity flow pressure 0.35 bar
Usable bath opening
(W x L / D)
16 x 3 / 14 cm
Filling volume 2 ... 3 l
Dimensions
(W × L × H)
24 x 46 x 40 cm

Product Features

  • Small overall dimensions for lowest space requirement
  • Splash-proof keypad
  • Large, bright LED display
  • Precise PID temperature control
  • RS232 interface for PC connection
  • Classification 3 according to DIN12876-1
  • Pump connections for external temperature applications
  • Bath opening for temperature control of small objects in the bath
  • Front drain
Dimensions and volumes
Dimensions
(W × L × H)
24 x 46 x 40 cm
Filling volume 2 ... 3 l
Weight 35 kg
Pump connections M16x1  male
Barbed fittings inner diameter 8/12 mm
Temperature values
Working temperature range -30 ... +150 °C
Temperature stability ±0.03 °C
Temperature display resolution 0.1 °C
Ambient temperature +5 ... +40 °C
Data based on voltage version:
230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F)
Performance values
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20
0.32 0.25 0.15
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20
kW 0.32 0.25 0.15
Pump capacity flow rate 15 l/min
Pump capacity flow pressure 0.35 bar
Viscosity max. 30 cST
Refrigerant stage 1
Refrigerant R134a
Global Warming Potential for R134a 1430
Filling volume 150 g
Carbon dioxide equivalent 0.215 t
Performance values
Heating capacity 1 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20
0.32 0.25 0.15
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20
kW 0.32 0.25 0.15
Pump capacity flow rate 15 l/min
Pump capacity flow pressure 0.35 bar
Viscosity max. 30 cST
Refrigerant stage 1
Refrigerant R134a
Global Warming Potential for R134a 1430
Filling volume 150 g
Carbon dioxide equivalent 0.215 t
Performance values
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20
0.32 0.25 0.15
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20
kW 0.32 0.25 0.15
Pump capacity flow rate 15 l/min
Pump capacity flow pressure 0.35 bar
Viscosity max. 30 cST
Refrigerant stage 1
Refrigerant R134a
Global Warming Potential for R134a 1430
Filling volume 150 g
Carbon dioxide equivalent 0.215 t
Performance values
Heating capacity 2 kW
Cooling capacity
(Medium: Ethanol)
°C
kW
20 0 -20
0.32 0.25 0.15
Cooling capacity
(Medium: Ethanol)
°C 20 0 -20
kW 0.32 0.25 0.15
Pump capacity flow rate 15 l/min
Pump capacity flow pressure 0.35 bar
Viscosity max. 30 cST
Refrigerant stage 1
Refrigerant R134a
Global Warming Potential for R134a 1430
Filling volume 150 g
Carbon dioxide equivalent 0.215 t
Bath
Bath tank Stainless steel
Bath cover integrated
Usable bath opening
(W x L / D)
16 x 3 / 14 cm
Cooling
Cooling of compressor 1-stage Air
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump type Immersion Pump
Electronics
Temperature control PID1
Temperature display LED
Available voltage versions
230V/60Hz Nema N6-20 Plug 9400330.13
115V/60Hz Nema N5-20 Plug 9400330.02
230V/50Hz Schuko Plug - CEE 7/4 Plug Type F 9400330.03
230V/50Hz UK Plug Type BS1363A 9400330.04
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)

Accessory for CF30

We would also be happy to send you our printed catalogs and brochures by post.


Request brochure

Benefits

  • PID2
    Precise
    PID Temperature control with set control parameters, temperature stability ±0.02...±0.2 °C
  • Made in Germany
    JULABO. Quality.
    Highest standards of quality for a long product life.
  • Blatt wächst in einer Glühbirne
    Green technology.
    Development consistently applied environmentally friendly materials and technologies.
  • network
    Satisfied customers.
    11 subsidiaries and more than 100 partners worldwide guarantee fast and qualified JULABO support.