FT900 FT Immersion Cooler

with immersion probe for rapid cooling of liquids

JULABO Immersion Coolers are primarily used for counter-cooling when connected to a heating circulator, or for rapidly cooling fluids down to low temperatures. Take advantage of the space-saving design.

JULABO Immersion Coolers represent an economical alternative to conventional tap water cooling. The easy-to-use units are suitable for a variety of cooling applications.

Request a quote
Data based on voltage version:
230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F)
FT900
Order No. 9 650 890.05
Working temperature range -90 ... +30 °C
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -20 -40 -80
0.3 0.27 0.27 0.24 0.2 0.07
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -20 -40 -80
kW 0.3 0.27 0.27 0.24 0.2 0.07
Dimensions
(W × L × H)
38 x 55 x 60 cm
FT900
Order No. 9 650 890.05
Working temperature range -90 ... +30 °C
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -20 -40 -80
0.3 0.27 0.27 0.24 0.2 0.07
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -20 -40 -80
kW 0.3 0.27 0.27 0.24 0.2 0.07
Dimensions
(W × L × H)
38 x 55 x 60 cm
FT900
Order No. 9 650 890.02
Working temperature range -90 ... +30 °C
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -20 -40 -80
0.3 0.27 0.27 0.24 0.2 0.07
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -20 -40 -80
kW 0.3 0.27 0.27 0.24 0.2 0.07
Dimensions
(W × L × H)
38 x 55 x 60 cm
FT900
Order No. 9 650 890.03
Working temperature range -90 ... +30 °C
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -20 -40 -80
0.3 0.27 0.27 0.24 0.2 0.07
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -20 -40 -80
kW 0.3 0.27 0.27 0.24 0.2 0.07
Dimensions
(W × L × H)
38 x 55 x 60 cm
FT900
Order No. 9 650 890.03
Working temperature range -90 ... +30 °C
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -20 -40 -80
0.3 0.27 0.27 0.24 0.2 0.07
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -20 -40 -80
kW 0.3 0.27 0.27 0.24 0.2 0.07
Dimensions
(W × L × H)
38 x 55 x 60 cm
FT900
Order No. 9 650 890.04
Working temperature range -90 ... +30 °C
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -20 -40 -80
0.3 0.27 0.27 0.24 0.2 0.07
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -20 -40 -80
kW 0.3 0.27 0.27 0.24 0.2 0.07
Dimensions
(W × L × H)
38 x 55 x 60 cm
FT900
Order No. 9 650 890.04
Working temperature range -90 ... +30 °C
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -20 -40 -80
0.3 0.27 0.27 0.24 0.2 0.07
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -20 -40 -80
kW 0.3 0.27 0.27 0.24 0.2 0.07
Dimensions
(W × L × H)
38 x 55 x 60 cm

Product Features

  • Rapid cooling of liquids down to low temperatures
  • Counter-cooling of liquids in combination with heating circulators
  • Dry ice substitution
  • Preventing the use of precious tap water for cooling
  • Compact design, small footprint
  • Ease of operation
Dimensions and volumes
Dimensions
(W × L × H)
38 x 55 x 60 cm
Weight 50 kg
Temperature values
Working temperature range -90 ... +30 °C
Ambient temperature 5 ... 35 °C
Data based on voltage version:
230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F)
Performance values
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -20 -40 -80
0.3 0.27 0.27 0.24 0.2 0.07
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -20 -40 -80
kW 0.3 0.27 0.27 0.24 0.2 0.07
Refrigerant stage 1
Refrigerant R404A
Global Warming Potential for R404A 3922
Filling volume 310 g
Carbon dioxide equivalent 1.216 t
Refrigerant stage 2
Refrigerant R23
Global Warming Potential for R23 14800
Filling volume 65 g
Carbon dioxide equivalent 0.962 t
Performance values
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -20 -40 -80
0.3 0.27 0.27 0.24 0.2 0.07
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -20 -40 -80
kW 0.3 0.27 0.27 0.24 0.2 0.07
Refrigerant stage 1
Refrigerant R404A
Global Warming Potential for R404A 3922
Filling volume 310 g
Carbon dioxide equivalent 1.216 t
Refrigerant stage 2
Refrigerant R23
Global Warming Potential for R23 14800
Filling volume 65 g
Carbon dioxide equivalent 0.962 t
Performance values
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -20 -40 -80
0.3 0.27 0.27 0.24 0.2 0.07
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -20 -40 -80
kW 0.3 0.27 0.27 0.24 0.2 0.07
Refrigerant stage 1
Refrigerant R404A
Global Warming Potential for R404A 3922
Filling volume 310 g
Carbon dioxide equivalent 1.216 t
Refrigerant stage 2
Refrigerant R23
Global Warming Potential for R23 14800
Filling volume 65 g
Carbon dioxide equivalent 0.962 t
Performance values
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -20 -40 -80
0.3 0.27 0.27 0.24 0.2 0.07
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -20 -40 -80
kW 0.3 0.27 0.27 0.24 0.2 0.07
Refrigerant stage 1
Refrigerant R404A
Global Warming Potential for R404A 3922
Filling volume 310 g
Carbon dioxide equivalent 1.216 t
Refrigerant stage 2
Refrigerant R23
Global Warming Potential for R23 14800
Filling volume 65 g
Carbon dioxide equivalent 0.962 t
Performance values
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -20 -40 -80
0.3 0.27 0.27 0.24 0.2 0.07
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -20 -40 -80
kW 0.3 0.27 0.27 0.24 0.2 0.07
Refrigerant stage 1
Refrigerant R404A
Global Warming Potential for R404A 3922
Filling volume 310 g
Carbon dioxide equivalent 1.216 t
Refrigerant stage 2
Refrigerant R23
Global Warming Potential for R23 14800
Filling volume 65 g
Carbon dioxide equivalent 0.962 t
Performance values
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -20 -40 -80
0.3 0.27 0.27 0.24 0.2 0.07
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -20 -40 -80
kW 0.3 0.27 0.27 0.24 0.2 0.07
Refrigerant stage 1
Refrigerant R404A
Global Warming Potential for R404A 3922
Filling volume 310 g
Carbon dioxide equivalent 1.216 t
Refrigerant stage 2
Refrigerant R23
Global Warming Potential for R23 14800
Filling volume 65 g
Carbon dioxide equivalent 0.962 t
Performance values
Cooling capacity
(Medium: Ethanol)
°C
kW
20 10 0 -20 -40 -80
0.3 0.27 0.27 0.24 0.2 0.07
Cooling capacity
(Medium: Ethanol)
°C 20 10 0 -20 -40 -80
kW 0.3 0.27 0.27 0.24 0.2 0.07
Refrigerant stage 1
Refrigerant R404A
Global Warming Potential for R404A 3922
Filling volume 310 g
Carbon dioxide equivalent 1.216 t
Refrigerant stage 2
Refrigerant R23
Global Warming Potential for R23 14800
Filling volume 65 g
Carbon dioxide equivalent 0.962 t
Cooling
Cooling of compressor 2-stage Air
Other
IP Code
according to IEC 60529
IP 21
Immersion cooler
Dimensions probe tube
(L x ⌀)
65 x 1.5 cm
Immersion probe type flexible
Available voltage versions
230V/50-60Hz CH Plug Type SEV 1011 9650890.05
115V/60Hz Nema N5-15 Plug 9650890.02
230V/50-60Hz Schuko Plug - CEE 7/4 Plug Type F 9650890.03
230V/50-60Hz UK Plug Type BS1363A 9650890.04

Accessory for FT900

Brochure

Instruments for special requirements
English
PDF - 1.4 MB

Download

Operation Manual

Betriebsanleitung FD, FT Lemosa
English
PDF - 4.32 MB

Download

Product Data Sheet

FT900
English
PDF - 202.47 KB

Download

We would also be happy to send you our printed catalogs and brochures by post.

Request brochure

Benefits

  • Made in Germany
    JULABO. Quality.
    Highest standards of quality for a long product life.
  • Blatt wächst in einer Glühbirne
    Green technology.
    Development consistently applied environmentally friendly materials and technologies.
  • network
    Satisfied customers.
    11 subsidiaries and more than 100 partners worldwide guarantee fast and qualified JULABO support.
  • 100% geprüft
    100% Checked.
    100% testing. 100% quality. Each JULABO Circulator undergoes thorough quality testing before leaving the factory.
  • Quickstart
    Quick start.
    Individual JULABO consultation and comprehensive manuals at your disposal.
  • Downloads
    Services 24/7.
    Around the clock availability. You can find suitable accessories, data sheets, manuals, case studies, and more at www.julabo.com.