Case Study: PRESTO W80 Process system
Cooling a Rettberg 5 l reactor from +20 °C to -60 °C with a PRESTO W80 Process system
Objective
This case study tests the cooling capacity of the JULABO PRESTO W80 Process system with a 5 l glass reactor. The PRESTO W80 is connected to the reactor via two 1 in. long metal hoses. The PRESTO W80 Process system has been programmed to cool down from +20 °C to -60 °C.
Test Conditions
| Product | PRESTO W80 Process system | ||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||||||||||
| Heating capacity | 1.8 kW | ||||||||||||||||||||||
| Pump pressure | 5.8 psi | Band limit | °k | ||||||||||||||||||||
| Duration | 1:10 hr:min | Test volume | 5 l | ||||||||||||||||||||
| Setpoint 1 | +20 °C | Setpoint 2 | -60 °C | ||||||||||||||||||||
| Reactor | 5l Rettberg (glass) | Tubing length | 1 in. | ||||||||||||||||||||
| Fluid in unit | Thermal HL80 | Fluid in reactor | Thermal HL80 | ||||||||||||||||||||
Environment
| Room temperature | 20 °C |
|---|---|
| Humidity | 45 % |
| Voltage version | 230V/50Hz |