PRESTO W80 Process system

2-stage process circulators for low temperatures

As air- or water-cooled versions, the units with 2-stage refrigeration unit offer high cooling and heating capacities for lowest temperatures down to -80 °C.

The highly dynamic temperature control systems PRESTO are designed for high-precision temperature control for a wide range of applications such as reactor vessels or material stress tests. Moreover, by using efficient components, the process systems can compensate exothermic and endothermic reactions exceptionally fast. Permanent internal monitoring and self-lubricating pumps ensure a long life-time. In addition, numerous interfaces offer many remote control possibilities across networks or for integration into higher-level control systems.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
W80
Order number including voltage version 9421801.14
Working temperature range -80 ... +250 °C
Temperature stability ±0.01 ... ±0.05 °C
Heating capacity 1.5 kW
Pump capacity flow pressure 0 ... 40 l/min
Pump capacity pressure 1.5 ... 18.9 psi
Total dimensions
(W × L × H)
16.9 x 25.6 x 49.6 in.
Arrange callback

Product Features

  • Analog connections, RS485, Profibus DP (accessory)
  • Space optimized design creates more room directly next to the units
  • External Pt100 sensor connection
  • Built-in 5.7’’ industrial color touchscreen
  • Second external Pt100 sensor connection (accessory)
  • Alarm output
  • Temperature stability ±0.01 °C ... ±0.05 °C
  • Pump pressure up to 1.7 bar, max. flow rate 40 l/min
  • Heating capacity up to 3.4 kW
  • Cooling capacity up to 1.2 kW

Technical data for PRESTO W80

Dimensions and volumes
Total dimensions
(W × L × H)
16.9 x 25.6 x 49.6 in.
Internal usable expansion volume 5.6 l
Minimal process volume 3.9 l
Active heat exchanger volume 1.7 l
Weight 350.5 lbs
Pump connections M24x1.5 male
Cooling Water Connection G¾ in.
Temperature values
Working temperature range -80 ... +250 °C
Temperature stability ±0.01 ... ±0.05 °C
Setting the resolution of the temperature display 0.01 °C
Temperature display resolution 0.01 °C
Ambient temperature +5 ... +40 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 1.5 kW
Cooling capacity
Medium: Ethanol
°C 200 100 20 0 -20 -30 -40 -60 -80
kW 1 1.2 1.2 1.2 1.2 1.1 1.1 1 0.65 0.1
Refrigerant Information Refrigerant stage 1 Refrigerant stage 2
Refrigerant R507 R23
Global Warming Potential 3985 14800
Filling weight 720 g 500 g
Carbon dioxide equivalent 2.8692 t 7.4 t

1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.

Pump capacity flow pressure 0 ... 40 l/min
Pump capacity pressure 1.5 ... 18.9 psi
Viscosity max. 50 cSt
Power consumption 15 A
Cooling
Cooling of compressor 2-stage Water
Cooling water pressure max. 87 psi
Max. heat dissipated by unit into cooling water 3 kW

Recommended cooling water properties
to provide full cooling capacity according to catalog values
Cooling water temperature range 10 ... 15 °C
Cooling water difference pressure 21.8 ... 87 psi
Cooling water consumption 1 2 l/min

1 Cooling water consumption may vary outside recommended cooling water properties.


Permissible cooling water properties
Cooling water temperature range 5 ... 35 °C
Cooling water difference pressure 2 7.3 ... 87 psi

2 At cooling water temperatures of 25 °C and higher, the minimum differential pressure is 1 bar.

Other
Sound pressure level 64 dbA
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 20
Pump type Centrifugal Pump (Magnetically coupled)
Electronics
Interfaces Alarm output / Ethernet / Modbus TCP/IP / Profibus optional / REG/EPROG optional / RS232 / RS485 optional / SD memory card / Standby-Input optional / USB
External pt100 sensor connection integrated
2nd external Pt100 sensor connection accessory
Integrated programmer 8x60 steps
Temperature control ICC
Absolute temperature calibration 3 Point Calibration
Temperature display 5.7" TFT Touchscreen
Temperature setting Touchscreen
Available voltage versions
9421801.03 230V 50Hz (Schuko Plug - CEE 7/4 Plug Type F) (R507)
9421801.04 230V 50Hz (UK Plug Type BS1363A) (R507)
9421801.14 208V 60Hz (Nema N5-20 Plug) (R507)
Contact expert

Cool-down time

Medium: Thermal HL

Heat-up time

Medium: Thermal HL

Pump capacity

Accessory for PRESTO W80

Downloads for PRESTO W80

We would also be happy to send you our printed catalogs and brochures by post.

Request brochure

PRESTO W80 Benefits

Touch display. Perfect operation.

With the touch display, the user always has an overview of all values and functions. The intuitive and multilingual menu structure enables perfect control.

Convenience for several users

Administrator level for customizing instrument settings, user levels with limited permissions for fast and safe defined access, password protection, all levels adjustable

100 % Cooling capacity

‘Active Cooling Control’ for cooling available throughout the entire working temperature range, fast cool-down even at higher temperatures

Intelligent temperature control.

Intelligent cascade control - automatic and self-optimizing adaptation of the PID control parameters with external stability of +/- 0.05 °C.

Show all 21 benefits