Case Study: PRESTO A30 Process system
Cooling a QVF 6 l reactor from +150 °C to +20 °C with a PRESTO A30 Process system
Objective
This case study tests the cooling capacity of the JULABO PRESTO A30 Process system with a 6 l glass reactor. The PRESTO A30 is connected to the reactor via two 2 in. long metal hoses. The PRESTO A30 Process system has been programmed to cool down from +150 °C to +20 °C.
![Case study Presto A30 6-liter reactor](https://dbyt75tph3sip.cloudfront.net/image/512030579289/image_l0bp46s75908j7jtlhm9jp5251/-S421-FJPG/casestudy-presto-a30-qvf-6-liters-reactor.jpg)
Test Conditions
Product | PRESTO A30 Process system |
---|---|
Heating capacity | 2.7 kW |
Pump pressure | 7.3 psi |
Duration | 1:35 hr:min |
Test volume | 5 l |
Setpoint 1 | +150 °C |
Setpoint 2 | +20 °C |
Reactor | 6l QVF (glass) |
Tubing length | 2 in. |
Fluid in unit | Thermal HL60 |
Fluid in reactor | Thermal HL60 |
Environment
Room temperature | 20 °C |
---|---|
Humidity | 45 % |
Voltage version | 230V/50Hz |