FT400 FT Immersion Cooler

with immersion probe for rapid cooling of liquids

JULABO Immersion Coolers are primarily used for counter-cooling when connected to a heating circulator, or for rapidly cooling fluids down to low temperatures. Take advantage of the space-saving design.

JULABO Immersion Coolers represent an economical alternative to conventional tap water cooling. The easy-to-use units are suitable for a variety of cooling applications.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
FT400
Order No. 9650840.02
Working temperature range -40 ... +30 °C
Cooling capacity
Medium: Ethanol
°C 20 10 0 -20 -40
kW* 0.45
0.36
0.3
0.14
0.03
Refrigerant R449A
Global Warming Potential 1397
Filling volume 112 g
Carbon dioxide equivalent 0.156 t
* Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
Dimensions
(W × L × H)
7.9 x 11.8 x 16.9 in.
Arrange callback Accessory

Product Features

  • Rapid cooling of liquids down to low temperatures
  • Counter-cooling of liquids in combination with heating circulators
  • Preventing the use of precious tap water for cooling
  • Compact design, small footprint
  • Ease of operation
  • Dry ice substitution
Dimensions and volumes
Dimensions
(W × L × H)
7.9 x 11.8 x 16.9 in.
Weight 52.9 lbs
Temperature values
Working temperature range -40 ... +30 °C
Ambient temperature +5 ... +35 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Cooling capacity
Medium: Ethanol
°C 20 10 0 -20 -40
kW* 0.45
0.36
0.3
0.14
0.03
Refrigerant R449A
Global Warming Potential 1397
Filling volume 112 g
Carbon dioxide equivalent 0.156 t
* Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
Power 4 A
Cooling
Cooling of compressor 1-stage Air
Other
Sound pressure level 61 dbA
IP Code
according to IEC 60529
IP 21
Immersion cooler
Immersion probe type inflexible
Dimensions immersion probe cylindrical
(L x ⌀)
4.7 x 2 in.
Available voltage versions
9650840.02 115V/60Hz (Nema N5-15 Plug) (R449A)
9650840.03 230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F) (R452A)
9650840.04 230V/50Hz (UK Plug Type BS1363A) (R452A)
9650840.05 230V/50Hz (CH Plug Type SEV 1011) (R452A)

Accessory for FT400

Contact expert

We would also be happy to send you our printed catalogs and brochures by post.


Request brochure

Benefits

  • 100% geprüft
    100% Checked.
    100% testing. 100% quality. Each JULABO Circulator undergoes thorough quality testing before leaving the factory.
  • Blatt wächst in einer Glühbirne
    Green technology.
    Development consistently applied environmentally friendly materials and technologies.
  • Made in Germany
    JULABO. Quality.
    Highest standards of quality for a long product life.
  • Quickstart
    Quick start.
    Individual JULABO consultation and comprehensive manuals at your disposal.