FK31-SL Calibration Bath

with integrated cooling machine for working temperatures up to -30 °C

The calibration baths in this category feature an integrated refrigeration unit and are suitable for calibration applications to -30 °C. The bath cover with openings and Viton® sleeves is already included in delivery for all models.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
FK31-SL
Order No. 9352628.02
Working temperature range -30 ... +200 °C
Temperature stability ±0.005 °C
Heating capacity 1 kW
Cooling capacity
Medium: Ethanol
°C 20 0 -20
kW* 0.46
0.34
0.15
Refrigerant R134a
Global Warming Potential 1430
Filling volume 200 g
Carbon dioxide equivalent 0.286 t
* Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 5.8 ... 10.2 psi
Usable bath opening
(W x L / D)
4.7 in.
Filling volume 24 l
Dimensions
(W × L × H)
12.6 x 17.7 x 35.8 in.
Arrange callback Accessory

Product Features

  • LCD DIALOG DISPLAY backlit for convenient interactive operation
  • ICC (Intelligent Cascade Control), self-optimizing temperature control
  • TCF Temperature Control Features to optimize the control behavior
  • Integrated programmer for 6 x 60 program steps
  • Highest temperature stability to ±0.005 °C (±0.02 °C at +200 °C)
  • Temperature uniformity better than ±0.01 °C
  • Optional Pt100 sensor for reference temperature measurements
  • Temperature chamber with overflow at maximum liquid level
Dimensions and volumes
Dimensions
(W × L × H)
12.6 x 17.7 x 35.8 in.
Filling volume 24 l
Usable immersion depth 12.2 in.
Weight 112.4 lbs
Temperature values
Working temperature range -30 ... +200 °C
Temperature stability ±0.005 °C
Temperature display resolution 0.01 °C
Ambient temperature +5 ... +40 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 1 kW
Cooling capacity
Medium: Ethanol
°C 20 0 -20
kW* 0.46
0.34
0.15
Refrigerant R134a
Global Warming Potential 1430
Filling volume 200 g
Carbon dioxide equivalent 0.286 t
* Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 5.8 ... 10.2 psi
Maximum suction 2.9 ... 5.8 psi
Viscosity max. 70 cST
Power 7 A
Bath
Bath tank Stainless steel
Bath cover integrated
Usable bath opening
(W x L / D)
4.7 in.
Cooling
Cooling coil not available
Cooling of compressor 1-stage Air
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump type Immersion Pump
Electronics
Interfaces Profibus optional
External pt100 sensor connection integrated
Integrated programmer 6x60 steps
Temperature control ICC
Absolute temperature calibration 3 Point Calibration
Temperature displayTemperature display VFD
Temperature settingTemperature setting Keypad
Available voltage versions
9352628.02 115V/60Hz (Nema N5-15 Plug) (R134a)
9352628.03 230V/50Hz (Schuko Plug - CEE 7/4 Plug Type F) (R134a)
9352628.04 230V/50Hz (UK Plug Type BS1363A) (R134a)
Included in delivery
Including bath cover with seven openings.

Accessory for FK31-SL

Contact expert

We would also be happy to send you our printed catalogs and brochures by post.


Request brochure

Benefits

  • 100% geprüft
    100% Checked.
    100% testing. 100% quality. Each JULABO Circulator undergoes thorough quality testing before leaving the factory.
  • Blatt wächst in einer Glühbirne
    Green technology.
    Development consistently applied environmentally friendly materials and technologies.
  • Made in Germany
    JULABO. Quality.
    Highest standards of quality for a long product life.
  • Quickstart
    Quick start.
    Individual JULABO consultation and comprehensive manuals at your disposal.