Case Study: PRESTO W50 Process system
Heating up of an QVF 50 l reactor from -40 °C to +20 °C with a PRESTO W50 Process system
Objective
This case study tests the heating capacity of the JULABO PRESTO W50 Process system with a 50 l glass reactor. The PRESTO W50 is connected to the reactor via two 2 in. metal hoses. The PRESTO W50 Process system has been programmed to heat up from -40 °C to +20 °C.
Test Conditions
| Product | PRESTO W50 Process system | ||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||||
| Heating capacity | 6 kW | ||||||||||||||||
| Pump pressure | 7.3 psi | Band limit | °k | ||||||||||||||
| Duration | 55 hr:min | Test volume | 35 l | ||||||||||||||
| Setpoint 1 | -40 °C | Setpoint 2 | +20 °C | ||||||||||||||
| Reactor | 50l QVF (glass) | Tubing length | 2 in. | ||||||||||||||
| Fluid in unit | Thermal HL60 | Fluid in reactor | Thermal HL60 | ||||||||||||||
Environment
| Room temperature | 20 °C |
|---|---|
| Humidity | 45 % |
| Voltage version | 400V/3PNPE/50Hz |
Control Parameters
| XP | 0.2 K |
| TN | 695 s |
| TV | 85 s |
| XPU | 15 K |