Case Study: PRESTO W50t Process system
Heating up of an Büchiglas 50 l reactor from -20 °C to +20 °C with a PRESTO W50t Process system
Objective
This case study tests the heating capacity of the JULABO PRESTO W50t Process system with a 50 l glass reactor. The PRESTO W50t is connected to the reactor via two 2 in. metal hoses. The PRESTO W50t Process system has been programmed to heat up from -20 °C to +20 °C.
Test Conditions
| Product | PRESTO W50t Process system | ||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||||||
| Heating capacity | 12 kW | ||||||||||||||||||
| Pump pressure | 7.3 psi | Band limit | °k | ||||||||||||||||
| Duration | 33 hr:min | Test volume | 37.5 l | ||||||||||||||||
| Setpoint 1 | -20 °C | Setpoint 2 | +20 °C | ||||||||||||||||
| Reactor | 50l Büchiglas (glass) | Tubing length | 2 in. | ||||||||||||||||
| Fluid in unit | Thermal HL60 | Fluid in reactor | Thermal HL60 | ||||||||||||||||
Environment
| Room temperature | 20 °C |
|---|---|
| Humidity | 45 % |
| Voltage version | 400V/3PNPE/50Hz |
Control Parameters
| XP | 0.19 K |
| TN | 695 s |
| TV | 85 s |
| XPU | 15 K |
Test Results
The PRESTO W50t Process system heats up the reactor from -20 °C to +20 °C in 33 min.
- Set point
- Temperature in the application
- Temperature in temperature control unit