Case Study: PRESTO W91 Process system
Cooling a QVF 50 l reactor from +20 °C to lowest possible temperature with a PRESTO W91 Process system
Objective
This case study tests the lowest temperature of the JULABO PRESTO W91 Process system with a 50 l glass reactor. The PRESTO W91 is connected to the reactor via two 2 in. metal hoses. The PRESTO W91 Process system cools from +20 °C to -88 °C maximum low temperature.
Test Conditions
| Product | PRESTO W91 Process system | ||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||||
| Heating capacity | 18 kW | ||||||||||||||||
| Pump pressure | 7.3 psi | ||||||||||||||||
| Duration | 5:10 hr:min | ||||||||||||||||
| Test volume | 35 l | ||||||||||||||||
| Setpoint 1 | +20 °C | ||||||||||||||||
| Setpoint 2 | -95 °C | ||||||||||||||||
| Reactor | 50l QVF (glass) | ||||||||||||||||
| Tubing length | 2 in. | ||||||||||||||||
| Fluid in unit | Thermal HL80 | ||||||||||||||||
| Fluid in reactor | Thermal HL80 | ||||||||||||||||
Environment
| Room temperature | 20 °C |
|---|---|
| Humidity | 45 % |
| Voltage version | 400V/3PNPE/50Hz |