Case Study: PRESTO W91 Process system
Cooling a QVF 50 l reactor from +20 °C to lowest possible temperature with a PRESTO W91 Process system
Objective
This case study tests the lowest temperature of the JULABO PRESTO W91 Process system with a 50 l glass reactor. The PRESTO W91 is connected to the reactor via two 2 m metal hoses. The PRESTO W91 Process system cools from +20 °C to -88 °C maximum low temperature.
![Process system PRESTO W91with QVF 50 liter reactor](https://dbyt75tph3sip.cloudfront.net/image/512030579289/image_cdffigdnth3mfctl7ki8s9op1k/-S421-FJPG/casestudy-presto-w91-qvf-50-liters-reactor.jpg)
Test Conditions
Product | PRESTO W91 Process system | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||||
Heating capacity | 18 kW | ||||||||||||||||
Pump pressure | 0.5 bar | ||||||||||||||||
Duration | 5:10 hr:min | ||||||||||||||||
Test volume | 35 l | ||||||||||||||||
Setpoint 1 | +20 °C | ||||||||||||||||
Setpoint 2 | -95 °C | ||||||||||||||||
Reactor | 50l QVF (glass) | ||||||||||||||||
Tubing length | 2 m | ||||||||||||||||
Fluid in unit | Thermal HL80 | ||||||||||||||||
Fluid in reactor | Thermal HL80 |
Environment
Room temperature | 20 °C |
---|---|
Humidity | 45 % |
Voltage version | 400V/3PNPE/50Hz |