Case Study: PRESTO W56 Process system
Cooling a Büchiglas 50 l reactor from +20 °C to -20 °C with a PRESTO W56 Process system
Objective
This case study tests the cooling capacity of the JULABO PRESTO W56 Process system with a 50 l glass reactor. The PRESTO W56 is connected to the reactor via two 2 in. long metal hoses. The PRESTO W56 Process system has been programmed to cool down from +20 °C to -20 °C.
Test Conditions
Product | PRESTO W56 Process system | ||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||||
Heating capacity | 27 kW | ||||||||||||||||
Band limit | 200 °k | ||||||||||||||||
Pump pressure | 7.3 psi | ||||||||||||||||
Duration | 43 hr:min | ||||||||||||||||
Test volume | 37.5 l | ||||||||||||||||
Setpoint 1 | +20 °C | ||||||||||||||||
Setpoint 2 | -20 °C | ||||||||||||||||
Reactor | 50l Büchiglas (glass) | ||||||||||||||||
Tubing length | 2 in. | ||||||||||||||||
Fluid in unit | Thermal HL60 | ||||||||||||||||
Fluid in reactor | Thermal HL60 |
Environment
Room temperature | 20 °C |
---|---|
Humidity | 45 % |
Voltage version | 400V/3PPE/50Hz |
Control Parameters
XP | 0.2 K |
TN | 1571 s |
TV | 93 s |
XPU | 55.94 K |
Test Results
The PRESTO W56 Process system cools the reactor from +20 °C to -20 °C in 43 min.
Set point |
Temperature in the application |
Temperature in temperature control unit |