Case Study: PRESTO W92tt Process system
Cooling a Büchiglas 100 l reactor from +20 °C to -50 °C with a PRESTO W92tt Process system
Objective
This case study tests the cooling capacity of the JULABO PRESTO W92tt Process system with a 100 l glass reactor. The PRESTO W92tt is connected to the reactor via two 3 in. long metal hoses. The PRESTO W92tt Process system has been programmed to cool down from +20 °C to -50 °C.
Test Conditions
| Product | PRESTO W92tt Process system | ||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||||||||||
| Heating capacity | 36 kW | ||||||||||||||||||||||
| Pump pressure | 7.3 psi | Band limit | °k | ||||||||||||||||||||
| Duration | 1:50 hr:min | Test volume | 70 l | ||||||||||||||||||||
| Setpoint 1 | +20 °C | Setpoint 2 | -50 °C | ||||||||||||||||||||
| Reactor | 100l Büchiglas (glass) | Tubing length | 3 in. | ||||||||||||||||||||
| Fluid in unit | Thermal HL80 | Fluid in reactor | Thermal HL80 | ||||||||||||||||||||
Environment
| Room temperature | 20 °C |
|---|---|
| Humidity | 45 % |
| Voltage version | 400V/3PNPE/50Hz |