Case Study: MAGIO MX-2500F Refrigerated / heating circulator
Cooling a Asahi 20 l reactor from +20 °C to -20 °C with a MAGIO MX-2500F Refrigerated / heating circulator
Objective
This case study tests the cooling capacity of the JULABO MAGIO MX-2500F Refrigerated / heating circulator with a 20 l glass reactor. The MAGIO MX-2500F is connected to the reactor via two 1 in. long metal hoses. The MAGIO MX-2500F Refrigerated / heating circulator has been programmed to cool down from +20 °C to -20 °C.
Test Conditions
Product | MAGIO MX-2500F Refrigerated / heating circulator | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||
Heating capacity | 3 kW | ||||||||||||
Pump pressure | 7.3 psi | ||||||||||||
Duration | 49 hr:min | ||||||||||||
Test volume | 15 l | ||||||||||||
Setpoint 1 | +20 °C | ||||||||||||
Setpoint 2 | -20 °C | ||||||||||||
Reactor | 20l Asahi (glass) | ||||||||||||
Tubing length | 1 in. | ||||||||||||
Fluid in unit | Thermal H10 | ||||||||||||
Fluid in reactor | Thermal H10 |
Environment
Room temperature | 20 °C |
---|---|
Humidity | 45 % |
Voltage version | 230V/50Hz |
Control Parameters
XP | 0.11 K |
TN | 200 s |
TV | 10 s |
XPU | 3.6 K |
Test Results
The MAGIO MX-2500F Refrigerated / heating circulator cools the reactor from +20 °C to -20 °C in 49 min.
Set point |
Temperature in the application |
Temperature in temperature control unit |