Cooling a Asahi 20 l reactor from +20 °C to -20 °C with a MAGIO MX-2500F Refrigerated / heating circulator
Objective
This case study tests the cooling capacity of the JULABO MAGIO MX-2500F Refrigerated / heating circulator with a 20 l glass reactor. The MAGIO MX-2500F is connected to the reactor via two 1 m long metal hoses. The MAGIO MX-2500F Refrigerated / heating circulator has been programmed to cool down from +20 °C to -20 °C.
Test Conditions
| Product | MAGIO MX-2500F Refrigerated / heating circulator | ||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||
| Heating capacity | 3 kW | ||||||||||||||
| Pump pressure | 0.5 bar | Band limit | °k | ||||||||||||
| Duration | 49 hr:min | Test volume | 15 l | ||||||||||||
| Setpoint 1 | +20 °C | Setpoint 2 | -20 °C | ||||||||||||
| Reactor | 20l Asahi (glass) | Tubing length | 1 m | ||||||||||||
| Fluid in unit | Thermal H10 | Fluid in reactor | Thermal H10 | ||||||||||||
Environment
| Room temperature | 20 °C |
|---|---|
| Humidity | 45 % |
| Voltage version | 230V/50Hz |
Control Parameters
| XP | 0.11 K |
| TN | 200 s |
| TV | 10 s |
| XPU | 3.6 K |
Test Results
The MAGIO MX-2500F Refrigerated / heating circulator cools the reactor from +20 °C to -20 °C in 49 min.
- Set point
- Temperature in the application
- Temperature in temperature control unit