CF40 Cryo-Compact Circulator
The first choice for routine and straightforward laboratory applications from -40°C to 150°C
Cryo-Compact Circulator for internal and external temperature applications
The CF series offers compact refrigerated/heating circulators. With small overall dimensions the units can even be placed on a very small surface or within a complex installation. All models offer a heating capacity of 2 kW and classification 3 according to DIN12876-1. The maximum ambient temperature of +40°C as well as a ventilation air cooling allow installing the units very closely side by side with other units or directly under a fume hood. Furthermore all the models feature pump connections for external temperature applications and a bath opening for temperature control of small objects directly in the circulator bath. Wetted parts are made of high quality stainless steel or plastic. The instruments provide a reliable Microprocessor electronics with high temperature stability as well as warning and safety functions.
Your advantages
- Ergonomic design and easy operation
- Small overall dimensions for lowest space requirement
- Splash-proof keypad
- Large, bright LED temperature display
- Precise PID temperature control
- RS232 interface for PC connection
- Adjustable high temperature cut-off
- Classification 3 according to DIN12876-1
- Pump connections for external temperature applications
- Bath opening for temperature control of small objects in the bath
- Front drain
Technical Data
| Order No. | 9400340 | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| Model series | Economy | ||||||||
| Category | Cryo-Compact Circulators | ||||||||
| Working temperature range (°C) | -40 ... +150 | ||||||||
| Temperature stability (°C) | ±0.03 | ||||||||
| Setting / display resolution | 0.1 °C | ||||||||
| Temperature Display | LED | ||||||||
| Heating capacity (kW) | 1 | ||||||||
| Cooling capacity (Medium Ethanol) |
|
||||||||
| Pump capacity flow rate (l/min) | 15 | ||||||||
| Pump capacity flow pressure (psi) | 5.1 | ||||||||
| Bath opening / bath depth (W x L / D inch) | 7.5 x 1.2 / 7.5 | ||||||||
| Pump connections | M16x1 | ||||||||
| Barbed fittings diameter (inner dia. / mm) | 8 / 12 | ||||||||
| Filling volume (liters) | 5.5 | ||||||||
| Refrigerant stage 1 | R404A | ||||||||
| Filling volume refrigerant stage 1 (g) | 153 | ||||||||
| Global Warming Potential for | 3922 | ||||||||
| Carbon dioxide equivalent stage 1 (t) | 0.6 | ||||||||
| Digital interfaces |
RS232 Optional Profibus |
||||||||
| Ambient temperature | 5...40 °C | ||||||||
| Dimensions W x L x H (inch) | 11 x 18.1 x 18.1 | ||||||||
| Weight (LBS) | 90 | ||||||||
| Classification according to DIN12876-1 | Classification III (FL) | ||||||||
| Included with each unit | 2 each barbed fittings for tubing 8 and 12 mm inner dia. (pump connections M16x1 female). | ||||||||
| Cooling of compressor | Air | ||||||||
| Power requirement V / Hz / A | 115/60/16 Additional voltage versions available |
Characteristics
Large LED temperature display for actual value and setpoint (resolution 0.1 °C)
Comfortable and simple operation for setpoint adjustment
PID Temperature control with set control parameters, temperature stability ±0.02...±0.2 °C
RS232 interface for PC connection, e.g. for data communication and recording of measured values
Adjustable high temperature cut-off or dry-running protection
Classification III (FL) according to DIN 12876-1
Technical changes without prior notification.
Images may deviate from the original.
We would be pleased to submit a non-binding offer for the selected product or answer all pertinent questions.
click here for quotation request CF40Search
Contact
JULABO USA Inc.
884 Marcon Boulevard
Allentown, PA 18109
Phone: 610-231-0250
Phone: 800-458-5226
(Toll-Free)
sales@julabo.us
julabo.us
Contact form
Available voltage versions
Order No.: 9400340
230 V / 50 Hz
230 V / 60 Hz
115 V / 60 Hz