FPW52-SL Ultra-Low Refrigerated / Heating Circulator

low temperatures below -50 °C for internal and external applications

The Ultra-Low Refrigerated Circulators of the HighTech Series with HL or SL circulator feature powerful pressure and suction pumps. The instruments provide the entire variety of functions of the professional HighTech Series of circulators.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
FPW52-SL
Order No. 9352753.16
Working temperature range -55 ... +100 °C
Temperature stability ±0.05 °C
Heating capacity 3 kW
Cooling capacity
Medium: Ethanol
°C 20 0 -20 -40 -50
kW* 3
2.8
1.6
0.65
0.26
Refrigerant R452A
Global Warming Potential 2140
Filling volume 850 g
Carbon dioxide equivalent 1.819 t
* Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 5.8 ... 10.2 psi
Usable bath opening
(W x L / D)
11 x 9.1 / 8.7 in.
Filling volume 17 ... 24 l
Dimensions
(W × L × H)
23.2 x 29.9 x 45.7 in.
Arrange callback Accessory

Product Features

  • LCD DIALOG DISPLAY backlit for convenient interactive operation
  • ICC (Intelligent Cascade Control), self-optimizing temperature control
  • TCF Temperature Control Features to optimize the control behavior
  • RS232/RS485 interface for online communication
  • Integrated programmer for 6 x 60 program steps
  • Connections for solenoid valve and HSP booster pump
  • Proportional cooling control
Dimensions and volumes
Dimensions
(W × L × H)
23.2 x 29.9 x 45.7 in.
Filling volume 17 ... 24 l
Weight 337.3 lbs
Pump connections M16x1 male
Cooling Water Connection G¾ in.
Temperature values
Working temperature range -55 ... +100 °C
Temperature stability ±0.05 °C
Setting the resolution of the temperature display 0.01 °C
Temperature display resolution 0.01 °C
Ambient temperature +5 ... +40 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 3 kW
Cooling capacity
Medium: Ethanol
°C 20 0 -20 -40 -50
kW* 3
2.8
1.6
0.65
0.26
Refrigerant R452A
Global Warming Potential 2140
Filling volume 850 g
Carbon dioxide equivalent 1.819 t
* Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
Pump capacity flow rate 22 ... 26 l/min
Pump capacity flow pressure 5.8 ... 10.2 psi
Maximum suction 2.9 ... 5.8 psi
Viscosity max. 70 cST
Power 24 A
Bath
Bath cover integrated
Usable bath opening
(W x L / D)
11 x 9.1 / 8.7 in.
Cooling
Cooling of compressor 1-stage Water
Cooling water pressure max. 87 psi
Cooling water difference pressure 50.8 psi
Cooling water temperature max. 20 °C
1.0Cooling water consumption 5 l/min
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Electronics
Interfaces Profibus optional
External pt100 sensor connection integrated
Integrated programmer 6x60 steps
Temperature control ICC
Absolute temperature calibration 3 Point Calibration
Temperature displayTemperature display VFD
Temperature settingTemperature setting Keypad
Available voltage versions
9352753.16 208V/3PPE/60Hz (Without Plug) (R452A)
9352753.07 400V/3PNPE/50Hz (Plug 32A CEE) (R452A)
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)

Accessory for FPW52-SL

Contact expert

We would also be happy to send you our printed catalogs and brochures by post.


Request brochure

Benefits

  • 100% geprüft
    100% Checked.
    100% testing. 100% quality. Each JULABO Circulator undergoes thorough quality testing before leaving the factory.
  • Blatt wächst in einer Glühbirne
    Green technology.
    Development consistently applied environmentally friendly materials and technologies.
  • Made in Germany
    JULABO. Quality.
    Highest standards of quality for a long product life.
  • Quickstart
    Quick start.
    Individual JULABO consultation and comprehensive manuals at your disposal.