CORIO CP-1000FW Refrigerated / heating circulatornew with an enhanced pump performance and a temperature range up to +200 °C
Performance data depends on the mains voltage in the country of use. Please select:
Country of use | Mains voltage | Plug | JULABO voltage version code
Order No. | 9013727.02 | |||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -50 ... +200 °C | |||||||||||||||||||||||||||
Temperature stability | ±0.03 °C | |||||||||||||||||||||||||||
Heating capacity | 1 kW | |||||||||||||||||||||||||||
Cooling capacity
* Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
|
||||||||||||||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | |||||||||||||||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | |||||||||||||||||||||||||||
Usable bath opening (W x L / D) |
7.1 x 5.1 / 5.9 in. | |||||||||||||||||||||||||||
Filling volume | 5 ... 7.5 l | |||||||||||||||||||||||||||
Dimensions (W × L × H) |
16.5 x 19.3 x 29.1 in. |
Product Features
- Models for internal and external applications
- Bright, white, easy to read display
- Very quiet
- Easy pump change-over between internal and external circulation
- External pump connections
- Powerful and infinitely adjustable pressure pump
- USB connection
- RS232 interface for online communication
- Bath lid and drain tap included
- Removable ventilation grid
- Refrigeration unit without side vents
- Unique early warning system for low liquid level
- Class III (FL) according to DIN 12876-1
- Space-saving cooling coil design provides more usable space in the bath tank
Dimensions (W × L × H) |
16.5 x 19.3 x 29.1 in. |
---|---|
Filling volume | 5 ... 7.5 l |
Weight | 113.5 lbs |
Pump connections | M16x1 male |
Barbed fittings inner diameter | 8/12 mm |
Working temperature range | -50 ... +200 °C |
---|---|
Temperature stability | ±0.03 °C |
Temperature display resolution | 0.01 ... 0.1 °C |
Ambient temperature | +5 ... +40 °C |
Performance data depends on the mains voltage in the country of use. Please select:
Country of use | Mains voltage | Plug | JULABO voltage version code
Heating capacity | 1 kW | ||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity
* Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
|
|||||||||||||||||||||||||||||
Pump capacity flow rate | 8 ... 27 l/min | ||||||||||||||||||||||||||||
Pump capacity flow pressure | 1.5 ... 10.2 psi | ||||||||||||||||||||||||||||
Viscosity max. | 50 cST | ||||||||||||||||||||||||||||
Power | 16 A |
Bath tank | Stainless steel |
---|---|
Bath cover | integrated |
Usable bath opening (W x L / D) |
7.1 x 5.1 / 5.9 in. |
Cooling of compressor | 1-stage Water |
---|---|
Cooling water pressure max. | 87 psi |
Cooling water difference pressure | 4.4 psi |
Cooling water temperature max. | 30 °C |
Cooling water consumption | 1.1 l/min |
Classification according to DIN12876-1 |
Classification III (FL) |
---|---|
Pump function | Pressure Pump |
Pump type | Immersion Pump |
Interfaces | RS232 / USB |
---|---|
Temperature control | PID1 |
Absolute temperature calibration | 3 Point Calibration |
Temperature display | LED |
Temperature setting | Keypad |
Electronic Timer hr:min |
0 ... 999 |
9013727.02 | 115V/60Hz (Nema N5-20 Plug) (R449A) |
---|---|
9013727.04 | 200-230V/50-60Hz (UK Plug Type BS1363A) (R449A) |
9013727.05 | 200-230V/50-60Hz (CH Plug Type SEV 1011) (R449A) |
9013727.33 | 200-230V/50-60Hz (Schuko Plug - CEE 7/4 Plug Type F) (R449A) |
9013727.33.chn | 200-230V/50-60Hz (CN Plug) (R449A) |
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male) |
Overview
Contact expertAccessory for CORIO CP-1000FW
Benefits
-
Safety.
CORIO CD and CP comply with Class III (FL) according to DIN 12876-1 and switches off automatically in case of high temperature or low liquid level alarm. -
Highest measuring accuracy
‘Absolute Temperature Calibration’ for manual compensation of a temperature difference, 3-point calibration -
Quick support
If an error occurs, the integrated Black-Box function permits fast diagnosis by the JULABO service team -
Brilliant.
Very bright display makes it easy to read even from a distance.