Case Study: PRESTO A40 Process system

Heating up of an QVF 6 l reactor from +20 °C to +50 °C with a PRESTO A40 Process system

Objective

This case study tests the heating capacity of the JULABO PRESTO A40 Process system with a 6 l glass reactor. The PRESTO A40 is connected to the reactor via two 2 in. metal hoses. The PRESTO A40 Process system has been programmed to heat up from +20 °C to +50 °C.

Test Conditions

Product PRESTO A40 Process system
Heating capacity 2.7 kW
Pump pressure 7.3 psi Band limit °k
Duration 28 hr:min Test volume 5 l
Setpoint 1 +20 °C Setpoint 2 +50 °C
Reactor 6l QVF (glass) Tubing length 2 in.
Fluid in unit Thermal HL60 Fluid in reactor Thermal HL60

Environment

Room temperature 20 °C
Humidity 45 %
Voltage version 230V/50Hz
PRESTO A40 details

Test Results

The PRESTO A40 Process system heats up the reactor from +20 °C to +50 °C in 28 min.

  • Set point
  • Temperature in the application
  • Temperature in temperature control unit