Case Study: PRESTO A40 Process system
Heating up of an QVF 6 l reactor from -20 °C to -10 °C with a PRESTO A40 Process system
Objective
This case study tests the heating capacity of the JULABO PRESTO A40 Process system with a 6 l glass reactor. The PRESTO A40 is connected to the reactor via two 2 in. metal hoses. The PRESTO A40 Process system has been programmed to heat up from -20 °C to -10 °C.
Test Conditions
Product | PRESTO A40 Process system | ||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||
Heating capacity | 2.7 kW | ||||||||||||||
Pump pressure | 7.3 psi | ||||||||||||||
Duration | 55 hr:min | ||||||||||||||
Test volume | 5 l | ||||||||||||||
Setpoint 1 | -20 °C | ||||||||||||||
Setpoint 2 | -10 °C | ||||||||||||||
Reactor | 6l QVF (glass) | ||||||||||||||
Tubing length | 2 in. | ||||||||||||||
Fluid in unit | Thermal HL60 | ||||||||||||||
Fluid in reactor | Thermal HL60 |
Environment
Room temperature | 20 °C |
---|---|
Humidity | 45 % |
Voltage version | 230V/50Hz |
Test Results
The PRESTO A40 Process system heats up the reactor from -20 °C to -10 °C in 55 min.
Set point |
Temperature in the application |
Temperature in temperature control unit |