Case Study: PRESTO W80 Process system

Cooling a Rettberg 5 l reactor from +50 °C to +20 °C with a PRESTO W80 Process system

Objective

This case study tests the cooling capacity of the JULABO PRESTO W80 Process system with a 5 l glass reactor. The PRESTO W80 is connected to the reactor via two 1 in. long metal hoses. The PRESTO W80 Process system has been programmed to cool down from +50 °C to +20 °C.

Test Conditions

Product PRESTO W80 Process system
Cooling capacity (Ethanol)
°C 200 100 20 0 -20 -30 -40 -60 -80
kW 1.2 1.2 1.2 1.2 1.1 1.1 1.1 0.65 0.1
Heating capacity 1.8 kW
Pump pressure 5.8 psi Band limit °k
Duration 30 hr:min Test volume 5 l
Setpoint 1 +50 °C Setpoint 2 +20 °C
Reactor 5l Rettberg (glass) Tubing length 1 in.
Fluid in unit Thermal HL80 Fluid in reactor Thermal HL80

Environment

Room temperature 20 °C
Humidity 45 %
Voltage version 230V/50Hz
PRESTO W80 details

Test Results

The PRESTO W80 Process system cools the reactor from +50 °C to +20 °C in 30 min.

  • Set point
  • Temperature in the application
  • Temperature in temperature control unit