Case Study: PRESTO A40 Process system
Cooling a QVF 6 l reactor from +20 °C to lowest possible temperature with a PRESTO A40 Process system
Objective
This case study tests the lowest temperature of the JULABO PRESTO A40 Process system with a 6 l glass reactor. The PRESTO A40 is connected to the reactor via two 2 in. metal hoses. The PRESTO A40 Process system cools from +20 °C to -24 °C maximum low temperature.
Test Conditions
| Product | PRESTO A40 Process system |
|---|---|
| Heating capacity | 2.7 kW |
| Pump pressure | 7.3 psi |
| Duration | 4:00 hr:min |
| Test volume | 5 l |
| Setpoint 1 | +20 °C |
| Setpoint 2 | -50 °C |
| Reactor | 6l QVF (glass) |
| Tubing length | 2 in. |
| Fluid in unit | Thermal HL60 |
| Fluid in reactor | Thermal HL60 |
Environment
| Room temperature | 20 °C |
|---|---|
| Humidity | 45 % |
| Voltage version | 230V/50Hz |