Case Study: PRESTO W56 Process system

Cooling a Büchiglas 100 l reactor from +150 °C to +20 °C with a PRESTO W56 Process system

Objective

This case study tests the cooling capacity of the JULABO PRESTO W56 Process system with a 100 l glass reactor. The PRESTO W56 is connected to the reactor via two 2 in. long metal hoses. The PRESTO W56 Process system has been programmed to cool down from +150 °C to +20 °C.

Test Conditions

Product PRESTO W56 Process system
Cooling capacity (Ethanol)
°C 20 0 -10 -20 -30 -40 -50
kW 25.8 23.1 16.9 11.5 7.1 3.5 1.4
Heating capacity 27 kW
Pump pressure 7.3 psi Band limit °k
Duration 1:01 hr:min Test volume 75 l
Setpoint 1 +150 °C Setpoint 2 +20 °C
Reactor 100l Büchiglas (glass) Tubing length 2 in.
Fluid in unit Thermal HL60 Fluid in reactor Thermal H20S

Environment

Room temperature 20 °C
Humidity 45 %
Voltage version 400V/3PPE/50Hz

Control Parameters

XP 0.2 K
TN 695 s
TV 85 s
XPU 15 K
PRESTO W56 details

Test Results

The PRESTO W56 Process system cools the reactor from +150 °C to +20 °C in 1:01 h.

  • Set point
  • Temperature in the application
  • Temperature in temperature control unit