CORIO CP-1200FW Refrigerated / heating circulator

with an enhanced pump performance and a temperature range up to +200 °C

Refrigerated circulators from the CORIO CP range are suitable for applications with a temperature range up to +200°C. The enhanced pump performance ensures they are suitable for easy temperature control tasks in combination with external applications.

Despite its very compact design, the water-cooled Refrigerated / heating circulator CORIO CP-1200FW, which has a working temperature range of , delivers a powerful cooling capacity of 1.24 kW at 20°C. Water-cooled systems are normally more efficient than air-cooled systems, as water has a higher specific heat capacity than air. In addition, they introduce far less waste heat into the environment, which helps create a pleasant working climate.

This cooling machine works with natural, environmentally-friendly refrigerant and was developed with a focus on energy efficiency. This means significant savings up to 70% on the operating costs for numerous applications, which also means rapid amortization of the procurement cost. At the same time, the lower energy consumption positively contributes to climate protection.

Energy-efficient temperature control

You can really save money with our energy-efficient refrigerated circulators. Thanks to speed-controlled components, power consumption is significantly reduced. This reduces electricity costs by up to 70% over the entire product life cycle. By minimizing energy requirements, you also make a positive contribution to protecting the environment.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
CP-1200FW
Order number including voltage version 9013728.N1.04
Working temperature range -50 ... 200 °C
Temperature stability ±0.03 °C
Heating capacity 2 kW
Pump capacity flow rate 8 ... 27 l/min
Pump capacity pressure 0.1 ... 0.7 bar
Usable bath opening
(W x L / D)
18 x 13 / 15 cm
Filling volume 5 ... 7.5 l
Total dimensions
(W × L × H)
33 x 47 x 70 cm
Arrange callback

Product Features

  • Models for internal and external applications
  • Bright, white, easy to read display
  • Very quiet
  • powerful and infinitely adjustable pressure pump
  • USB connection
  • RS232 interface for online communication
  • Bath lid and drain tap included
  • removable ventilation grid
  • Refrigeration unit without side vents
  • Class III (FL) according to DIN 12876-1
  • easy switching between internal and external circulation
  • Integrated pump connection M16×1
  • Space-saving cooling coil design yields more usable space in the bath tank
  • early warning system for low liquid level

Technical data for CORIO CP-1200FW

Dimensions and volumes
Total dimensions
(W × L × H)
33 x 47 x 70 cm
Filling volume 5 ... 7.5 l
Weight 51.5 kg
Pump connections M16x1 male
Barbed fittings inner diameter 8/12 mm
Temperature values
Working temperature range -50 ... 200 °C
Temperature stability ±0.03 °C
Temperature display resolution 0.01 °C
Ambient temperature +5 ... +40 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 2 kW
Cooling capacity
Medium: Ethanol
°C 20 0 -20 -30 -40
kW 1 1.24 1.09 0.62 0.37 0.2
Refrigerant R1270
Global Warming Potential 2
Filling weight 85 g
Carbon dioxide equivalent 0.00017 t

1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.

Pump capacity flow rate 8 ... 27 l/min
Pump capacity pressure 0.1 ... 0.7 bar
Viscosity max. 50 cSt
Power consumption 10 A
Bath
Bath tank Stainless steel
Bath cover integrated
Usable bath opening
(W x L / D)
18 x 13 / 15 cm
Cooling
Cooling of compressor 1-stage Water
Cooling water pressure max. 6 bar

Recommended cooling water properties
to provide full cooling capacity according to catalog values
Cooling water temperature range 10 ... 15 °C
Cooling water difference pressure 3.5 ... 6 bar
Cooling water consumption 1.1 l/min

Permissible cooling water properties
Cooling water temperature range 5 ... 35 °C
Cooling water difference pressure 0.5 ... 6 bar
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump function Pressure Pump
Pump type Immersion Pump
Electronics
Interfaces RS232 / USB
Temperature control PID1
Absolute temperature calibration 3 Point Calibration
Temperature display LED
Temperature setting Keypad
Electronic Timer
h:min
0 ... 999
Available voltage versions
9013728.N1.33.chn 200-230V 50/60Hz (CN Plug) (R1270)
9013728.N1.33 200-230V 50/60Hz (Schuko Plug - CEE 7/4 Plug Type F) (R1270)
9013728.N1.05 200-230V 50/60Hz (CH Plug Type T12 / 10A) (R1270)
9013728.N1.04 200-230V 50/60Hz (UK Plug Type BS1363A) (R1270)
9013728.N1.22 100-115V 50/60Hz (Nema N5-20 Plug) (R1270)
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)
Contact expert
Medium: Water

Heat-up time

Medium: Thermal

Cool-down time

Medium: Ethanol

Pump capacity

Medium: Water

Accessory for CORIO CP-1200FW

Downloads for CORIO CP-1200FW

We would also be happy to send you our printed catalogs and brochures by post.

Request brochure

CORIO CP-1200FW Benefits

Brilliant.

Very bright display makes it easy to read even from a distance.

Connection. Easy.

Inclined pump connections (M16×1) facilitate the connection of applications. Each unit includes 2 barbed fittings of 8/12 mm diameter each.

Everything at the front.

All operating controls and safety functions are accessed easily and comfortably from the front.

Exact.

You can rely on it. PID1 control and 'Active Cooling Control' make the new CORIO precise and perfect.

Show all 10 benefits