SL-14K HighTech SL calibration bath
for working temperatures from +50 °C to +300 °C
JULABO calibration baths are ideal for high precision calibration of sensors, measuring instruments, thermometers, etc. The instruments are certified for application in DKD laboratories and according to the requirements specified by DIN EN ISO 9001:2000.
Order No. | 9352514.33 |
---|---|
Working temperature range | +50 ... +300 °C |
Temperature stability | ±0.005 °C |
Heating capacity | 3 kW |
Pump capacity flow rate | 22 ... 26 l/min |
Pump capacity flow pressure | 0.4 ... 0.7 bar |
Usable bath opening (W x L / D) |
12 cm |
Filling volume | 14 l |
Dimensions
(W × L × H) |
22 x 46 x 61 cm |
Product Features
- LCD DIALOG DISPLAY backlit for convenient interactive operation
- ICC (Intelligent Cascade Control), self-optimizing temperature control
- TCF Temperature Control Features to optimize the control behavior
- Integrated programmer for 6 x 60 program steps
- Highest temperature stability to ±0.005 °C (±0.02 °C at +200 °C)
- Temperature uniformity better than ±0.01 °C
- Optional Pt100 sensor for reference temperature measurements
- Temperature chamber with overflow at maximum liquid level
Technical data for SL-14K
Dimensions (W × L × H) |
22 x 46 x 61 cm |
---|---|
Filling volume | 14 l |
Usable immersion depth | 31 cm |
Weight | 20 kg |
Working temperature range | +50 ... +300 °C |
---|---|
Temperature stability | ±0.005 °C |
Setting the resolution of the temperature display | 0.01 °C |
Ambient temperature 1 | +5 ... +40 °C |
1 Depending on the ambient temperature and the bath fluid used, it may be necessary to use the cooling coil for temperature applications around +50 °C. |
Heating capacity | 3 kW |
---|---|
Pump capacity flow rate | 22 ... 26 l/min |
Pump capacity flow pressure | 0.4 ... 0.7 bar |
Maximum suction | 0.2 ... 0.4 bar |
Viscosity max. | 70 cSt |
Power consumption | 14 A |
Bath tank | Stainless steel |
---|---|
Bath cover | integrated |
Usable bath opening (W x L / D) |
12 cm |
Cooling coil | integrated |
---|
Classification according to DIN12876-1 |
Classification III (FL) |
---|---|
IP Code according to IEC 60529 |
IP 21 |
Pump type | Immersion Pump |
Interfaces | Profibus optional |
---|---|
External pt100 sensor connection | integrated |
Integrated programmer | 6x60 steps |
Temperature control | ICC |
Absolute temperature calibration | 3 Point Calibration |
Temperature display | VFD |
Temperature setting | Keypad |
9352514.33 | 230V/50-60Hz (Schuko Plug - CEE 7/4 Plug Type F) |
---|
Including bath cover with seven openings |
Accessory for SL-14K
Downloads for SL-14K
We would also be happy to send you our printed catalogs and brochures by post.
Request brochureSL-14K Benefits
For flammable bath fluid
Classification III (FL) according to DIN 12876-1
ATC3. Calibration.
‘Absolute Temperature Calibration’ for compensating a physically caused temperature difference, 3-point calibration.
Process. Under control.
Full regulation of the dynamics control, access to all important control parameters for individual process optimization.
100 % Cooling capacity
‘Active Cooling Control’ for cooling available throughout the entire working temperature range, fast cool-down even at higher temperatures