Case Study: PRESTO A80t Process system
Heating up of an Asahi 20 l reactor from 0 °C to +20 °C with a PRESTO A80t Process system
Objective
This case study tests the heating capacity of the JULABO PRESTO A80t Process system with a 20 l glass reactor. The PRESTO A80t is connected to the reactor via two 1 m metal hoses. The PRESTO A80t Process system has been programmed to heat up from 0 °C to +20 °C.
![PRESTO A80 process system with Asahi glass reactor](https://dbyt75tph3sip.cloudfront.net/image/512030579289/image_6p9hvse4cp4v11j7vgaunumc1a/-S421-FJPG/casestudy-presto-a80-asahi-20-liters-reactor.jpg)
Test Conditions
Product | PRESTO A80t Process system | ||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||||||||
Heating capacity | 2.8 kW | ||||||||||||||||||||
Pump pressure | 0.5 bar | ||||||||||||||||||||
Duration | 45 hr:min | ||||||||||||||||||||
Test volume | 19 l | ||||||||||||||||||||
Setpoint 1 | 0 °C | ||||||||||||||||||||
Setpoint 2 | +20 °C | ||||||||||||||||||||
Reactor | 20l Asahi (glass) | ||||||||||||||||||||
Tubing length | 1 m | ||||||||||||||||||||
Fluid in unit | Thermal HL80 | ||||||||||||||||||||
Fluid in reactor | Thermal HL80 |
Environment
Room temperature | 20 °C |
---|---|
Humidity | 45 % |
Voltage version | 208V/60Hz |