Case Study: PRESTO W50t Process system
Cooling a Büchiglas 50 l reactor from +20 °C to lowest possible temperature with a PRESTO W50t Process system
Objective
This case study tests the lowest temperature of the JULABO PRESTO W50t Process system with a 50 l glass reactor. The PRESTO W50t is connected to the reactor via two 2 m metal hoses. The PRESTO W50t Process system cools from +20 °C to -39.30 °C maximum low temperature.
Test Conditions
Product | PRESTO W50t Process system | ||||||||||||||||
Cooling capacity (Ethanol)
|
|||||||||||||||||
Heating capacity | 12 kW | ||||||||||||||||
Pump pressure | 0.5 bar | ||||||||||||||||
Duration | 3:40 hr:min | ||||||||||||||||
Test volume | 37.5 l | ||||||||||||||||
Setpoint 1 | +20 °C | ||||||||||||||||
Setpoint 2 | -50 °C | ||||||||||||||||
Reactor | 50l Büchiglas (glass) | ||||||||||||||||
Tubing length | 2 m | ||||||||||||||||
Fluid in unit | Thermal HL60 | ||||||||||||||||
Fluid in reactor | Thermal HL60 |
Environment
Room temperature | 20 °C |
Humidity | 45 % |
Voltage version | 400V/3PNPE/50Hz |
Control Parameters
XP | 0.19 K |
TN | 695 s |
TV | 85 s |
XPU | 15 K |
Test Results
The PRESTO W50t Process system cools the reactor from +20 °C to the lowest temperature of -39.30 °C in 3:40 h.
Set point |
Temperature in the application |
Temperature in temperature control unit |