Case Study: PRESTO W91tt Process system
Cooling a Büchiglas 100 l reactor from +20 °C to -50 °C with a PRESTO W91tt Process system
Objective
This case study tests the cooling capacity of the JULABO PRESTO W91tt Process system with a 100 l glass reactor. The PRESTO W91tt is connected to the reactor via two 2 m long metal hoses. The PRESTO W91tt Process system has been programmed to cool down from +20 °C to -50 °C.
Test Conditions
| Product | PRESTO W91tt Process system | ||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||||
| Heating capacity | 36 kW | ||||||||||||||||
| Pump pressure | 0.45 bar | ||||||||||||||||
| Duration | 2:30 hr:min | ||||||||||||||||
| Test volume | 100 l | ||||||||||||||||
| Setpoint 1 | +20 °C | ||||||||||||||||
| Setpoint 2 | -50 °C | ||||||||||||||||
| Reactor | 100l Büchiglas (glass) | ||||||||||||||||
| Tubing length | 2 m | ||||||||||||||||
| Fluid in unit | Thermal HL80 | ||||||||||||||||
| Fluid in reactor | Thermal HL80 | ||||||||||||||||
Environment
| Room temperature | 20 °C |
|---|---|
| Humidity | 45 % |
| Voltage version | 400V/3PNPE/50Hz |