Case Study: PRESTO A40 Process system
Cooling a Asahi 20 l reactor from +170 °C to +100 °C with a PRESTO A40 Process system
Objective
This case study tests the cooling capacity of the JULABO PRESTO A40 Process system with a 20 l glass reactor. The PRESTO A40 is connected to the reactor via two 2 m long metal hoses. The PRESTO A40 Process system has been programmed to cool down from +170 °C to +100 °C.
![PRESTO A40 process system with Asahi glass reactor](https://dbyt75tph3sip.cloudfront.net/image/512030579289/image_8k3k4pq80d3b39r5uu4cokm15s/-S421-FJPG/casestudy-presto-a40-asahi-20-liters-reactor.jpg)
Test Conditions
Product | PRESTO A40 Process system | ||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity (Ethanol)
|
|||||||||||||||
Heating capacity | 2.7 kW | ||||||||||||||
Pump pressure | 0.4 bar | ||||||||||||||
Duration | 1:05 hr:min | ||||||||||||||
Test volume | 18 l | ||||||||||||||
Setpoint 1 | +170 °C | ||||||||||||||
Setpoint 2 | +100 °C | ||||||||||||||
Reactor | 20l Asahi (glass) | ||||||||||||||
Tubing length | 2 m | ||||||||||||||
Fluid in unit | Thermal HL60 | ||||||||||||||
Fluid in reactor | Thermal HL60 |
Environment
Room temperature | 20 °C |
---|---|
Humidity | 45 % |
Voltage version | 230V/50Hz |