CORIO CD-201F Refrigerated / heating circulator

The new generation of laboratory circulators for routine work and standard tasks

Refrigerated / heating circulators of the CORIO CD series distinguish themselves with a great price-to-performance ratio. Durable and dependable, they are suitable for a wide range of standard and routine work in both laboratories and industrial environments.

The Refrigerated / heating circulator CORIO CD-201F , with a working temperature range of -20 ... 150 °C, offers a powerful cooling capacity of 0.22 kW at 20 °C despite its very compact design.

Thanks to its special design, the CORIO CD-201F fits under laboratory shelves and fume hoods with limited headroom. In addition, the bath opening is at an ergonomic working height. If headroom is limited or if a lot of work is done at the bath, the CORIO CD-201F is the ideal solution. In comparison, the CORIO CD-201F is 21 centimeters lower than the 200F, while the 200F is the right choice when table space is limited.

The cooling machine uses a natural, environmentally friendly refrigerant.

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
CD-201F
Order number including voltage version 9012702.N1.04
Working temperature range -20 ... 150 °C
Temperature stability ±0.03 °C
Heating capacity 2 kW
Pump capacity flow rate 15 l/min
Pump capacity pressure 0.35 bar
Usable bath opening
(W x L / D)
13 x 15 / 15 cm
Filling volume 3 ... 4 l
Total dimensions
(W × L × H)
44 x 41 x 44 cm
Arrange callback

Product Features

  • Models for internal and external applications
  • Bright, white, easy to read display
  • Very quiet
  • easy switching between internal and external circulation
  • Bath lid and drain tap included
  • removable ventilation grid
  • Refrigeration unit without side vents
  • NRTL certification for the U.S. and Canada: Available upon request

Technical data for CORIO CD-201F

Dimensions and volumes
Total dimensions
(W × L × H)
44 x 41 x 44 cm
Filling volume 3 ... 4 l
Weight 25 kg
Pump connections M16x1 male
Barbed fittings inner diameter 8/12 mm
Temperature values
Working temperature range -20 ... 150 °C
Temperature stability ±0.03 °C
Temperature display resolution 0.01 °C
Ambient temperature +5 ... +40 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 2 kW
Cooling capacity
Refrigerant R290
Global Warming Potential 3
Filling weight 30 g
Carbon dioxide equivalent 9.0E-5 t

1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.

Pump capacity flow rate 15 l/min
Pump capacity pressure 0.35 bar
Viscosity max. 50 cSt
Power consumption 12 A
Bath
Bath tank Stainless steel
Bath cover integrated
Usable bath opening
(W x L / D)
13 x 15 / 15 cm
Cooling
Cooling of compressor 1-stage Air
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump function Pressure Pump
Pump type Immersion Pump
Electronics
Interfaces USB
Temperature control PID1
Absolute temperature calibration 1 Point Calibration
Temperature display LED
Temperature setting Keypad
Available voltage versions
9012702.N1.01 100V 50/60Hz (Nema N5-15 Plug) (R290)
9012702.N1.02 115V 60Hz (Nema N5-15 Plug) (R290)
9012702.N1.03 230V 50Hz (Schuko Plug - CEE 7/4 Plug Type F) (R290)
9012702.N1.03.chn 230V 50Hz (CN Plug) (R290)
9012702.N1.04 230V 50Hz (UK Plug Type BS1363A) (R290)
9012702.N1.05 230V 50Hz (CH Plug Type T12 / 10A) (R290)
9012702.N1.13 230V 60Hz (Nema N6-20 Plug) (R290)
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)
Contact expert

Cool-down time

Medium: Thermal
Medium: Ethanol

Pump capacity

Medium: Water

Accessory for CORIO CD-201F

Downloads for CORIO CD-201F

We would also be happy to send you our printed catalogs and brochures by post.

Request brochure

CORIO CD-201F Benefits

Brilliant.

Very bright display makes it easy to read even from a distance.

Timer. Integrated.

CORIO circulators include an integrated timer function. When the set time has elapsed, a signal sounds and the device switches off. Setting range: 0 ... 999 minutes.

Connection. Easy.

Inclined pump connections (M16×1) facilitate the connection of applications. Each unit includes 2 barbed fittings of 8/12 mm diameter each.

Everything at the front.

All operating controls and safety functions are accessed easily and comfortably from the front.

Show all 28 benefits