CORIO CD-200F Refrigerated / heating circulator

The new generation of laboratory circulators for routine work and standard tasks

Refrigerated / heating circulators of the CORIO CD series distinguish themselves with a great price-to-performance ratio. Durable and dependable, they are suitable for a wide range of standard and routine work in both laboratories and industrial environments.

With a working temperature range of -20 ... +150 °C, the Refrigerated / heating circulator CORIO CD-200F delivers an outstanding cooling capacity of 0.22 kW at 20°C despite its compact design. The optimized design of the bath tank and cooling coil makes the bath more spacious and user-friendly for the operator. Ventilation slots and connections are located on the front and rear, allowing several units to be placed next to each other in order to save space.

With natural refrigerant

The CORIO CD-200F is available with natural refrigerant. Order No. 9012701.N1.03

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
CD-200F
Order number including voltage version see refrigerant variants
Working temperature range -20 ... +150 °C
Temperature stability ±0.03 °C
Heating capacity 2 kW
Pump capacity flow pressure 15 l/min
Pump capacity pressure 0.35 bar
Usable bath opening
(W x L / D)
13 x 15 / 15 cm
Filling volume 3 ... 4 l
Total dimensions
(W × L × H)
23 x 39 x 65 cm
Arrange callback

Product Features

  • Models for internal and external applications
  • Bright, white, easy to read display
  • Very quiet
  • USB connection
  • Integrated pump connection M16×1
  • easy switching between internal and external circulation
  • Bath lid and drain tap included
  • removable ventilation grid
  • Refrigeration unit without side vents
  • Class III (FL) according to DIN 12876-1
  • Space-saving cooling coil design yields more usable space in the bath tank

Technical data for CORIO CD-200F

Dimensions and volumes
Total dimensions
(W × L × H)
23 x 39 x 65 cm
Filling volume 3 ... 4 l
Weight 26 kg
Pump connections M16x1 male
Barbed fittings inner diameter 8/12 mm
Temperature values
Working temperature range -20 ... +150 °C
Temperature stability ±0.03 °C
Setting the resolution of the temperature display 0.1 °C
Temperature display resolution 0.01 ... 0.1 °C
Ambient temperature +5 ... +40 °C

Performance data depends on the mains voltage in the country of use. Please select:

Country of use | Mains voltage | Plug | JULABO voltage version code
Performance values
Heating capacity 2 kW
Cooling capacity
Refrigerant variant
Order number including voltage version 9012701.04
Medium: Ethanol
°C 20 0 -10 -20
kW 1 0.22 0.17 0.13 0.06
Refrigerant R134a
Global Warming Potential 1430
Filling weight 70 g
Carbon dioxide equivalent 0.1001 t
Refrigerant variant
Order number including voltage version 9012701.N1.04
Medium: Ethanol
°C 20 0 -10 -20
kW 1 0.22 0.17 0.13 0.06
Refrigerant R290
Global Warming Potential 3
Filling weight 26 g
Carbon dioxide equivalent 7.8E-5 t

1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.

Pump capacity flow pressure 15 l/min
Pump capacity pressure 0.35 bar
Viscosity max. 50 cSt
Power consumption 12 A
Bath
Bath tank Stainless steel
Bath cover integrated
Usable bath opening
(W x L / D)
13 x 15 / 15 cm
Cooling
Cooling of compressor 1-stage Air
Other
Classification
according to DIN12876-1
Classification III (FL)
IP Code
according to IEC 60529
IP 21
Pump function Pressure Pump
Pump type Immersion Pump
Electronics
Interfaces USB
Temperature control PID1
Absolute temperature calibration 1 Point Calibration
Temperature display LED
Temperature setting Keypad
Available voltage versions
9012701.13 230V 60Hz (Nema N6-20 Plug) (R134a)
9012701.N1.13 230V 60Hz (Nema N6-20 Plug) (R290)
9012701.05 230V 50Hz (CH Plug Type SEV 1011) (R134a)
9012701.N1.05 230V 50Hz (CH Plug Type SEV 1011) (R290)
9012701.04 230V 50Hz (UK Plug Type BS1363A) (R134a)
9012701.N1.04 230V 50Hz (UK Plug Type BS1363A) (R290)
9012701.03 230V 50Hz (Schuko Plug - CEE 7/4 Plug Type F) (R134a)
9012701.N1.03 230V 50Hz (Schuko Plug - CEE 7/4 Plug Type F) (R290)
9012701.03.chn 230V 50Hz (CN Plug) (R134a)
9012701.N1.03.chn 230V 50Hz (CN Plug) (R290)
9012701.02 115V 60Hz (Nema N5-15 Plug) (R134a)
9012701.N1.02 115V 60Hz (Nema N5-15 Plug) (R290)
9012701.01 100V 50/60Hz (Nema N5-15 Plug) (R134a)
9012701.N1.01 100V 50/60Hz (Nema N5-15 Plug) (R290)
Included in delivery
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male)
Contact expert

Cool-down time

Medium: Ethanol

Heat-up time

Medium: Thermal

Pump capacity

Medium: Water

Accessory for CORIO CD-200F

Downloads for CORIO CD-200F

We would also be happy to send you our printed catalogs and brochures by post.

Request brochure

CORIO CD-200F Benefits

Brilliant.

Very bright display makes it easy to read even from a distance.

Timer. Integrated.

CORIO circulators include an integrated timer function. When the set time has elapsed, a signal sounds and the device switches off. Setting range: 0 ... 999 minutes.

Connection. Easy.

Inclined pump connections (M16×1) facilitate the connection of applications. Each unit includes 2 barbed fittings of 8/12 mm diameter each.

Everything at the front.

All operating controls and safety functions are accessed easily and comfortably from the front.

Show all 9 benefits