Thermal HL30

JULABO Thermal HL30 is a water-glycol based bath fluid for closed systems and has excellent thermal properties.

Working temperature: -30 °C ... +90 °C

Benefits:

  • High stability
  • Low viscosity
  • Excellent heat conductivity
  • Minimum odor
  • Long fluid life

JULABO Thermal bath fluids are carefully selected and long-term tested media with ideal thermodynamic properties for a variety of applications. These bath fluids are ideally suited for temperature applications using JULABO instruments, and ensure safe, reliable operation as well as optimum heat transfer efficiency.

JULABO Thermal bath fluids are water-glycol based (monoethyleneglycol with anti-corrosion additives) and have excellent thermal properties and low viscosity. In addition, these bath fluids provide anti-freeze protection, meaning that they can be used at temperatures below the freezing point of water. Use the indicated mixing ratio! When using water-glycol based bath fluids for longer periods of time, regularly check the ratio of water to glycol in the mixture (50:50). If the proportion of glycol is too high, the fluid may become flammable. Conversely, if the proportion of water is too high, the fluid may freeze at low temperatures.

Please note: If the fluid is used below the specified lowest working temperature, this can cause the temperature to fluctuate or the overtemperature alarm to trigger because the fluid on the heater is too hot. Recommendation: Reduce the heater capacity during the heat-up process until the lowest specified working temperature is reached.

Technical data

Order number 5 liters 8940139
Order number 10 liters 8940138
Working temperature -30 ... 90 °C
Viscosity
(kinematic at 20 °C)
4.07 mm²/s
Density
(at 20 °C)
1.08 g / cm³
Pour point -70 °C
Boiling point +108 °C
Ignition temperature +430 °C
Color light yellow
Thermal expansion coefficient 0.0007 (g/(ml × K))
Heat conductivity 0.153 (W/(m × K))
Specific heat capacity 3.34 (kJ/(kg x K)@20°C)
Specific volume resistivity 200 (Ohm × cm)

Density

Specific heat capacity

Viscosity