Case Study: PRESTO A30 Process system
Heating up of an QVF 6 l reactor from +20 °C to +150 °C with a PRESTO A30 Process system
Objective
This case study tests the heating capacity of the JULABO PRESTO A30 Process system with a 6 l glass reactor. The PRESTO A30 is connected to the reactor via two 2 m metal hoses. The PRESTO A30 Process system has been programmed to heat up from +20 °C to +150 °C.
Test Conditions
| Product | PRESTO A30 Process system |
|---|---|
| Heating capacity | 2.7 kW |
| Pump pressure | 0.5 bar |
| Duration | 1:15 hr:min |
| Test volume | 5 l |
| Setpoint 1 | +20 °C |
| Setpoint 2 | +150 °C |
| Reactor | 6l QVF (glass) |
| Tubing length | 2 m |
| Fluid in unit | Thermal HL60 |
| Fluid in reactor | Thermal HL60 |
Environment
| Room temperature | 20 °C |
|---|---|
| Humidity | 45 % |
| Voltage version | 230V/50Hz |