FP55-SL Ultra-Low Refrigerated / Heating Circulator low temperatures below -50 °C for internal and external applications
Performance data depends on the mains voltage in the country of use. Please select:
Country of use | Mains voltage | Plug | JULABO voltage version code
Order No. | 9352755.16 | |||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Working temperature range | -55 ... +100 °C | |||||||||||||||||||||||
Temperature stability | ±0.05 °C | |||||||||||||||||||||||
Heating capacity | 3 kW | |||||||||||||||||||||||
Cooling capacity
* Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
|
||||||||||||||||||||||||
Pump capacity flow rate | 22 ... 26 l/min | |||||||||||||||||||||||
Pump capacity flow pressure | 5.8 ... 10.2 psi | |||||||||||||||||||||||
Usable bath opening (W x L / D) |
11 x 9.1 / 9.1 in. | |||||||||||||||||||||||
Filling volume | 27 l | |||||||||||||||||||||||
Dimensions (W × L × H) |
33.5 x 29.9 x 45.7 in. |
Product Features
- LCD DIALOG DISPLAY backlit for convenient interactive operation
- ICC (Intelligent Cascade Control), self-optimizing temperature control
- TCF Temperature Control Features to optimize the control behavior
- RS232/RS485 interface for online communication
- Integrated programmer for 6 x 60 program steps
- Connections for solenoid valve and HSP booster pump
- Proportional cooling control
Dimensions (W × L × H) |
33.5 x 29.9 x 45.7 in. |
---|---|
Filling volume | 27 l |
Weight | 396.8 lbs |
Pump connections | M16x1 male |
Working temperature range | -55 ... +100 °C |
---|---|
Temperature stability | ±0.05 °C |
Setting the resolution of the temperature display | 0.01 °C |
Temperature display resolution | 0.01 °C |
Ambient temperature | +5 ... +40 °C |
Performance data depends on the mains voltage in the country of use. Please select:
Country of use | Mains voltage | Plug | JULABO voltage version code
Heating capacity | 3 kW | ||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity
* Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids.
|
|||||||||||||||||||||||||
Pump capacity flow rate | 22 ... 26 l/min | ||||||||||||||||||||||||
Pump capacity flow pressure | 5.8 ... 10.2 psi | ||||||||||||||||||||||||
Maximum suction | 2.9 ... 5.8 psi | ||||||||||||||||||||||||
Viscosity max. | 70 cST | ||||||||||||||||||||||||
Power | 28 A |
Bath cover | integrated |
---|---|
Usable bath opening (W x L / D) |
11 x 9.1 / 9.1 in. |
Cooling of compressor | 1-stage Air |
---|
Classification according to DIN12876-1 |
Classification III (FL) |
---|---|
IP Code according to IEC 60529 |
IP 21 |
Interfaces | Profibus optional |
---|---|
External pt100 sensor connection | integrated |
Integrated programmer | 6x60 steps |
Temperature control | ICC |
Absolute temperature calibration | 3 Point Calibration |
Temperature displayTemperature display | VFD |
Temperature settingTemperature setting | Keypad |
9352755.07 | 400V/3PNPE/50Hz (Plug 32A CEE) (R452A) |
---|---|
9352755.S1.07 | 400V/3PNPE/50Hz (Plug 32A CEE) (R449A) |
9352755.16 | 208V/3PPE/60Hz (Without Plug) (R452A) |
2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male) |
Accessory for FP55-SL
Benefits
-
100% Checked.
100% testing. 100% quality. Each JULABO Circulator undergoes thorough quality testing before leaving the factory. -
Green technology.
Development consistently applied environmentally friendly materials and technologies. -
JULABO. Quality.
Highest standards of quality for a long product life. -
Quick start.
Individual JULABO consultation and comprehensive manuals at your disposal.