CORIO CP-310F Refrigerated / heating circulator
with an enhanced pump performance and a temperature range up to +200 °C
Refrigerated circulators from the CORIO CP range are suitable for applications with a temperature range up to +200°C. The enhanced pump performance ensures they are suitable for easy temperature control tasks in combination with external applications.
Performance data depends on the mains voltage in the country of use. Please select:
| Order number including voltage version | 9013713.N1.33 |
|---|---|
| Working temperature range | -30 ... 200 °C |
| Temperature stability | ±0.03 °C |
| Heating capacity | 1.5 kW |
| Pump capacity flow rate | 8 ... 27 l/min |
| Pump capacity pressure | 1.5 ... 11.6 psi |
|
Usable bath opening (W x L / D) |
5.1 x 5.9 / 5.9 in. |
| Filling volume | 3 ... 4 l |
|
Total dimensions
(W × L × H) |
9.1 x 15.7 x 25.6 in. |
Product Features
- Models for internal and external applications
- Bright, white, easy to read display
- Very quiet
- powerful and infinitely adjustable pressure pump
- Bath lid and drain tap included
- Removable ventilation grid
- Refrigeration unit without side vents
- easy switching between internal and external circulation
- early warning system for low liquid level
- NRTL-certified for the U.S. and Canada
Technical data for CORIO CP-310F
|
Total dimensions (W × L × H) |
9.1 x 15.7 x 25.6 in. |
|---|---|
| Filling volume | 3 ... 4 l |
| Weight | 56.4 lbs |
| Pump connections | M16x1 male |
| Barbed fittings inner diameter | 8/12 mm |
| Working temperature range | -30 ... 200 °C |
|---|---|
| Temperature stability | ±0.03 °C |
| Temperature display resolution | 0.01 °C |
| Ambient temperature | +5 ... +40 °C |
Performance data depends on the mains voltage in the country of use. Please select:
| Heating capacity | 1.5 kW | |||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Cooling capacity
1 Performance specifications measured in accordance with DIN 12876. Cooling capacities up to 20 °C measured with ethanol; over 20 °C with thermal oil unless otherwise specified. Performance specifications apply at an ambient temperature of 20 °C. Performance values may differ with other bath fluids. |
||||||||||||||||||||||
| Pump capacity flow rate | 8 ... 27 l/min | |||||||||||||||||||||
| Pump capacity pressure | 1.5 ... 11.6 psi | |||||||||||||||||||||
| Viscosity max. | 50 cSt | |||||||||||||||||||||
| Power consumption | 11 A | |||||||||||||||||||||
| Bath tank | Stainless steel |
|---|---|
| Bath cover | integrated |
|
Usable bath opening (W x L / D) |
5.1 x 5.9 / 5.9 in. |
| Cooling of compressor | 1-stage Air |
|---|
| Classification according to DIN12876-1 |
Classification III (FL) |
|---|---|
| IP Code according to IEC 60529 |
IP 21 |
| Pump function | Pressure Pump |
| Pump type | Immersion Pump |
| Interfaces | RS232 / USB |
|---|---|
| Temperature control | PID1 |
| Absolute temperature calibration | 3 Point Calibration |
| Temperature display | LED |
| Temperature setting | Keypad |
| Electronic Timer h:min |
0 ... 999 |
| 9013713.N1.01 | 100V 50/60Hz (Nema N5-15 Plug) (R290) |
|---|---|
| 9013713.N1.02 | 115V 60Hz (Nema N5-15 Plug) (R290) |
| 9013713.N1.33 | 200-230V 50/60Hz (Schuko Plug - CEE 7/4 Plug Type F) (R290) |
| 9013713.N1.33.chn | 200-230V 50/60Hz (CN Plug) (R290) |
| 9013713.N1.04 | 200-230V 50/60Hz (UK Plug Type BS1363A) (R290) |
| 9013713.N1.05 | 200-230V 50/60Hz (CH Plug Type T12 / 10A) (R290) |
| 2 Barbed fittings for tubing 8 and 12 mm ID. (Pump connections M16x1 male) |
Accessory for CORIO CP-310F
Downloads for CORIO CP-310F
We would also be happy to send you our printed catalogs and brochures by post.
Request brochureCORIO CP-310F Benefits
Brilliant.
Very bright display makes it easy to read even from a distance.
Connection. Easy.
Inclined pump connections (M16×1) facilitate the connection of applications. Each unit includes 2 barbed fittings of 8/12 mm diameter each.
Everything at the front.
All operating controls and safety functions are accessed easily and comfortably from the front.
Exact.
You can rely on it. PID1 control and 'Active Cooling Control' make the new CORIO precise and perfect.
Related Products