Case Study: PRESTO A40 Process system

Heating up of an QVF 6 l reactor from -25 °C to +200 °C with a PRESTO A40 Process system

Objective

This case study tests the heating capacity of the JULABO PRESTO A40 Process system with a 6 l glass reactor. The PRESTO A40 is connected to the reactor via two 2 in. metal hoses. The PRESTO A40 Process system has been programmed to heat up from -25 °C to +200 °C.

Test Conditions

Product PRESTO A40 Process system
Cooling capacity (Ethanol)
°C 20 0 -10 -20 -30 -40
kW 1.2 0.9 0.8 0.6 0.3 0.1
Heating capacity 2.7 kW
Pump pressure 5.8 psi Band limit °k
Duration 2:00 hr:min Test volume 10 l
Setpoint 1 -25 °C Setpoint 2 +200 °C
Reactor 6l QVF (glass) Tubing length 2 in.
Fluid in unit Thermal HL40 Fluid in reactor Thermal HL40

Environment

Room temperature 20 °C
Humidity 45 %
Voltage version 230V/50Hz
PRESTO A40 details

Test Results

The PRESTO A40 Process system heats up the reactor from -25 °C to +200 °C in 2:00 h.

  • Set point
  • Temperature in the application
  • Temperature in temperature control unit