Case Study: PRESTO A80t Process system

Heating up of an Asahi 20 l reactor from +20 °C to +50 °C with a PRESTO A80t Process system

Objective

This case study tests the heating capacity of the JULABO PRESTO A80t Process system with a 20 l glass reactor. The PRESTO A80t is connected to the reactor via two 1 in. metal hoses. The PRESTO A80t Process system has been programmed to heat up from +20 °C to +50 °C.

Test Conditions

Product PRESTO A80t Process system
Cooling capacity (Ethanol)
°C 200 100 20 0 -20 -30 -40 -60 -80
kW 1.2 1.2 1.2 1.2 1.1 1.1 1.1 0.65 0.1
Heating capacity 2.8 kW
Pump pressure 7.3 psi Band limit °k
Duration 55 hr:min Test volume 19 l
Setpoint 1 +20 °C Setpoint 2 +50 °C
Reactor 20l Asahi (glass) Tubing length 1 in.
Fluid in unit Thermal HL80 Fluid in reactor Thermal HL80

Environment

Room temperature 20 °C
Humidity 45 %
Voltage version 208V/60Hz
PRESTO A80t details

Test Results

The PRESTO A80t Process system heats up the reactor from +20 °C to +50 °C in 55 min.

  • Set point
  • Temperature in the application
  • Temperature in temperature control unit