Case Study: PRESTO A40 Process system
Cooling a Chemglas 10 l reactor from +100 °C to 0 °C with a PRESTO A40 Process system
Objective
This case study tests the cooling capacity of the JULABO PRESTO A40 Process system with a 10 l glass reactor. The PRESTO A40 is connected to the reactor via two 2 in. long metal hoses. The PRESTO A40 Process system has been programmed to cool down from +100 °C to 0 °C.
Test Conditions
| Product | PRESTO A40 Process system |
|---|---|
| Heating capacity | 2.7 kW |
| Pump pressure | 5.8 psi |
| Duration | 1:30 hr:min |
| Test volume | 10 l |
| Setpoint 1 | +100 °C |
| Setpoint 2 | 0 °C |
| Reactor | 10l Chemglas (glass) |
| Tubing length | 2 in. |
| Fluid in unit | Thermal HL40 |
| Fluid in reactor | Thermal HL40 |
Environment
| Room temperature | 20 °C |
|---|---|
| Humidity | 45 % |