Case Study: PRESTO A40 Process system

Cooling a Asahi 20 l reactor from +170 °C to -20 °C with a PRESTO A40 Process system

Objective

This case study tests the cooling capacity of the JULABO PRESTO A40 Process system with a 20 l glass reactor. The PRESTO A40 is connected to the reactor via two 2 in. long metal hoses. The PRESTO A40 Process system has been programmed to cool down from +170 °C to -20 °C.

Test Conditions

Product PRESTO A40 Process system
Heating capacity 2.7 kW
Pump pressure 5.8 psi Band limit °k
Duration 3:30 hr:min Test volume 18 l
Setpoint 1 +170 °C Setpoint 2 -20 °C
Reactor 20l Asahi (glass) Tubing length 2 in.
Fluid in unit Thermal HL60 Fluid in reactor Thermal HL60

Environment

Room temperature 20 °C
Humidity 45 %
Voltage version 230V/50Hz
PRESTO A40 details

Test Results

The PRESTO A40 Process system cools the reactor from +170 °C to -20 °C in 3:30 h.

  • Set point
  • Temperature in the application
  • Temperature in temperature control unit